• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30C07E20GYZ
  • Topic ID: id_13106859 Version: 3.0
  • Date: Aug 29, 2019 1:36:14 AM

Geometry Verification

This section contains a series of tests that allow determination of the proper x-y-z-gradient. Specific symptoms of this type of problem are images that display upside down, left-right reversed, offset scans that move in the opposite directions from those annotated on the images, or some combination of all of these problems. Run this series of tests in the order described to prevent confusing results. Once the correct polarities are established, the Signa system will make properly oriented head and body scans with any imaging database.

Note:

For a TwinSpeed scanner these tests should be repeated separately for each gradient mode, WHOLE-BODY and ZOOM.

Two Steps are required to insure proper gradient polarity