• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30862F20GYZ
  • Topic ID: id_13106982 Version: 3.0
  • Date: Aug 29, 2019 1:53:54 AM

1.5T HD 8 Channel Wrist Array MCQA Test Setup

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable15 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Cubical unified Phantom1-

5342681

-
Wrist array baseplate with baseplate riser1-

5160986-12

-
Table 3. Required conditions
ConditionReferenceEffectivity

The following coil configuration names must be installed to run this tool: Coil: HDWRIST If it is not installed, install that coil configuration with Config File Manager.

--

About this task

Follow this process to prepare for the MCQA test using the 1.5T HD 8 Channel Wrist Array by Invivo.

Procedure

  1. Place the baseplate of Wrist coil on the patient table as shown in Figure 1.
    Figure 1. Base Plate
  2. Attach the baseplate riser to the baseplate as shown in the Figure 2. Baseplate and the Baseplate riser in the locked position are shown in Figure 3.
    Figure 2. Base Plate and Base Plate Riser 1
    Figure 3. Base Plate and Base Plate Riser 2
  3. Attach 1.5T HD 8Ch Wrist Array to the baseplate assembly using the knobs is as shown in Figure 4. Slide the locked coil to centre and align properly.
    Figure 4. Attach 1.5T HD 8Ch Wrist Array
  4. Place the cubical unified phantom in the coil by lifting the upper portion of the coil as shown in the Figure 5. Close the coil and lock it (lock in Figure 6.
    Figure 5. Place the cubical unified phantom
  5. Connect the coil to respective port of the LPCA and landmark the coil at the marks shown using arrows as shown in Figure 6 and advance to scan.
    Figure 6. Lock and landmark
  6. Click here to go to the MCQA Test

Finalization

No finalization steps.