• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30867E20GYZ
  • Topic ID: id_13107077 Version: 3.0
  • Date: Aug 29, 2019 1:36:41 AM

LVShim Calibration File Generation Details

LVshim calibration files contain coil characterization data used by the LVshim analysis tool to calculate the new currents needed to correct for magnetic field inhomogeneity. Generic shim calibration files are supplied with the system software for all gradient coils and for GE S/C shim coils. Shim calibration files must be created at the site for Oxford and GE resistive shim coils. The generic gradient and S/C shim calibration files should work well at all sites, so it should not be necessary to create them. Creating a S/C shim calibration file is time-consuming and prone to errors. The procedures for creating a gradient or S/C shim calibration file on-site have not been supplied. Before creating a gradient or S/C shim calibration file at a site, contact your MAC Team Leader, or the OLC Magnet Support Group for assistance.

Note:

If the resistive shim coil assembly is replaced, or is moved within the magnet bore, then a new cal file may need to be created. Also, changes to system software may require the creation of a new cal file

Creating a Shim Calibration File

Note:

Creating a resistive shim calibration file can take two to four hours, and can be error-prone. Be sure there is adequate time to perform the task, and be careful during coil characterization to enter the correct perturbation currents.

The procedure in this section describes the creation of a 45-cm DSV Oxford Resistive Shim calibration file. The process of creating a GE Resistive Shim calibration file is similar. If you cannot collect image data to build a 45-cm DSV calibration file (due to phase wrap issues), then build a 32-cm DSV calibration file instead. Shim the magnet using the 32-cm DSV calibration file, but check the shim on a 45-cm DSV. You should be able to meet the 45-cm DSV shim specifications with the 32-cm DSV calibration file. If you cannot meet the 45-cm DSV shim specification, contact your local Zone Engineer or Support Center - Magnet Support Team for assistance.

A shim calibration file can be created in either of two ways:

  • Collect coil characterization scan data for all the coils in a coil set, then run the shim calibration processing on all the data to create the calibration matrix.

  • Collect coil characterization scan data for one coil in a coil set, then run the shim calibration processing on the data for that coil. This process is repeated until all the coils in the set have been characterized; then the calibration matrix is created.

Three Generation Options

There are three generation options: Create, Resume, and Modify. Table 1 shows all options.

Table 1. CAL FILE GENERATION OPTIONS
OPTIONCAL FILE TYPE
CREATEWhen this option is used, a new calibration file is created for the image data being analyzed. The new file contains the calibration file name token specified in menu item 4.
RESUMEThis option allows the user to resume a previously started shim calibration process. The software automatically displays the calibration file name token and sampling diameter of the latest partial shim file associated with the selected cal file type.
MODIFYThis option allows the modification of the data in an existing shim calibration file. The data is modified one coil at a time by re-scanning and reprocessing the baseline and perturbed states for a coil. The process is repeated for each coil to be re-characterized. The software automatically displays the calibration file name and sampling diameter of the latest shim file associated with the selected cal file type. To get a list of other shim calibration files associated with the selected cal file type, enter 4 at the Modify menu.

Suggested Perturb Currents and Series Descriptions

Perturbation Current

The Perturbation Current is the amount that the coil was perturbed during the collection of the selected image data set. If the perturbation value is set to 0 (zero), the selected image data set is treated as the baseline scan for the named coil.

This section contains the suggested perturb currents and series descriptions to be used during shim calibration file generation. The information in Table 2 to 9-coil gradient/resistive magnets. Table 3 applies to 12-coil resistive magnets.

Table 2. NINE-COIL GRADIENT/RESISTIVE (GE S-I, S-II, S-III)
SCAN INFORMATION
COIL NAMESCAN TYPESUGGESTED PERTURB CURRENTSERIES DESCRIPTION (SEE NOTE)
GRAD_XBaseline-XGRAD baseline
Perturbation50XGRAD 50
GRAD_YBaseline-YGRAD baseline
Perturbation50YGRAD 50
GRAD_ZBaseline-ZGRAD baseline
Perturbation50ZGRAD 50
Z2Baseline-Z2 baseline
Perturbation2.0Z2 2.0
Z3Baseline-Z3 baseline
Perturbation1.0Z3 1.0
ZXBaseline-ZX baseline
Perturbation1.0ZX 1.0
ZYBaseline-ZY baseline
Perturbation1.0ZY 1.0
X2-Y2Baseline-X2-Y2 baseline
Perturbation5.0X2-Y2 5.0
XYBaseline-XY baseline
Perturbation5.0XY 5.0
Note: 0 (zero) can be used in place of baseline for baseline scans. For example, XGRAD baseline is the same as XGRAD 0.
Table 3. TWELVE-COIL RESISTIVE (OXFORD)
SCAN INFORMATION
COIL NAMESCAN TYPESUGGESTED PERTURB CURRENTSERIES DESCRIPTION (SEE NOTE)
Z1Baseline-Z1 baseline
Perturbation0.1Z1 0.1
Z2Baseline-Z2 baseline
Perturbation0.5Z2 0.5
Z3Baseline-Z3 baseline
Perturbation2.0Z3 2.0
Z4Baseline Z4 baseline
Perturbation2.0Z4 2.0
XBaseline-X baseline
Perturbation0.1X 0.1
YBaseline Y baseline
Perturbation0.Y 0.1
ZXBaseline-ZX baseline
Perturbation1.0ZX 0.1
ZYBaseline-ZY baseline
Perturbation1.0ZY 1.0
XYBaseline-XY baseline
Perturbation1.0XY 1.0
X2-Y2Baseline-X2-Y2 baseline
Perturbation1.0X2-Y2 1.0
Z2XBaseline-Z2X baseline
Perturbation2.0Z2X 2.0
Z2YBaseline-Z2Y baseline
Perturbation2.0Z2Y 2.0
Note: 0 (zero) can be used in place of baseline for baseline scans. For example, Z1 baseline is the same as Z1 0

Enter Baseline Currents

Before starting data collection, enter the baseline currents for the coil set into the system. These are the final mechanical shim currents, or the currents used to get the best shim to date.

  1. At the operator workspace, select the tools icon in the desktop control panel.

  2. To run the LVShim tool from the Service Browser, follow the instructions on Figure 1, below:

Figure 1. STARTING THE LVSHIM TOOL FROM THE SERVICE BROWSER

For TwinSpeed, highlight the GradMode, then click on OK. Continue as shown in Table 4.

Table 4.
Output/PromptsINPUTS/COMMENTS


                    <<< LVshim Analysis >>>
(Magnet Type : Oxford (12 coil))

1. Shim Type (Gradient, Oxf-12,TEST): Gradient
2. Image Data (Ex, Se, Im Number)      : None Selected
3. Operation Mode (Service or Research): Service
4. Display Mag and Phase Diff Images   : No
5. Calibration File Name : lv_15t_3gc_22dsv.8645
6. Existing Current in Each Coil

XGRAD:  0.000   YGRAD:  0.000   ZGRAD: 0.000

Accept (q or s to quit).
Enter the Index Number to Change Default (0..6)[0]: ... 1 [Enter] To change Shim Type to
                                                         Resistive.
                  


                    <<< LVshim Analysis >>>
     (Magnet Type : Oxford (12 coil))
1. Shim Type (Gradient, Oxf-12,TEST): Oxf-12
2. Image Data (Ex, Se, Im Number)      : [None Selected]
3. Operation Mode (Service or Research): Service
4. Display Mag and Phase Diff Images   : No
5. Calibration File Name : lv_15t_12rc_45dsv.tag*
6. Existing Current in Each Coil.

    Z1:  0.000     X:  0.000      XY:  0.000
    Z2:  0.000     Y:  0.000   X2-Y2:  0.000
    Z3:  0.000    ZX:  0.000     Z2X:  0.000
    Z4:  0.000    ZY:  0.000     Z2Y:  0.000
0. Accept (q or s to quit).
Enter the Index Number to Change Default (0..6)[0]: ...         6 [Enter]

Coil 1 [ Z1 ] (-20.00..20.00) [0.0]: ..............   Enter baseline current for Coil 1 (Z1)
      <-------------------------------------          Enter baseline currents for Coils 2–11
Coil 12 [ Z2Y ] (-10.00..10.00) [0.0]: ............   Enter baseline current for Coil 12 (Z2Y)
<<< LVshim Analysis >>>
(Magnet Type : Oxford (12 coil))

1. Shim Type (Gradient, Oxf-12,TEST): Oxf-12
2. Image Data (Ex, Se, Im Number)      : [None Selected]
3. Operation Mode (Service or Research): Service
4. Display Mag and Phase Diff Images   : No
5. Calibration File Name : lv_15t_12rc_45dsv.tag*
6. Existing Current in Each Coil.

    Z1:  0.853     X: -3.407      XY:  0.310    
    Z2: -1.716     Y: -1.292   X2-Y2:  0.280    
    Z3:  1.489    ZX:  0.557     Z2X: -1.042    
    Z4:  4.112    ZY:  0.387     Z2Y:  0.641

0. Accept (q or s to quit).
Enter the Index Number to Change Default (0..6)[0]: ... 6 [Enter]
Coil 1 [ Z1 ] (-20.00..20.00) [0.0]: ..............      Enter baseline current for Coil 1 (Z1)
      <-------------------------------------             Enter baseline currents for Coils 2–11
Coil 12 [ Z2Y ] (-10.00..10.00) [0.0]: ............      Enter baseline current for Coil 12 (Z2Y)
<<< LVshim Analysis >>>
(Magnet Type : Oxford (12 coil))

1. Shim Type (Gradient, Oxf-12,TEST): Oxf-12
2. Image Data (Ex, Se, Im Number)      : [None Selected]
3. Operation Mode (Service or Research): Service
4. Display Mag and Phase Diff Images   : No
5. Calibration File Name : lv_15t_12rc_45dsv.tag*
6. Existing Current in Each Coil.

    Z1:  0.853     X: -3.407      XY:  0.310
    Z2: -1.716     Y: -1.292   X2-Y2:  0.280
    Z3:  1.489    ZX:  0.557     Z2X: -1.042
    Z4:  4.112    ZY:  0.387     Z2Y:  0.641

0. Accept (q or s to quit). ............... Make sure the correct currents are displayed, then:
Enter the Index Number to Change Default (0..6)[0]: ...   q [Enter] (To quit.)
Exit LVshim? (Y,N) [N] : ..............................   y [Enter]  (Exit LVshim.)
Press any key to quit --> .............................   Press any key
                  

Get Hard Copy of Perturbation Currents (Optional)

If a printer is available, and you wish to get a hard copy of the perturbation currents:

  1. At the operator workspace, select the tools icon in the desktop control panel.

  2. To run the LVShim tool from the Service Browser, follow the instructions on Figure 1

  3. For TwinSpeed, highlight the GradMode and click on OK. Continue as shown in Table 5.

Table 5. GETTING HARD COPY OF PERTURBATION CURRENTS (OPTIONAL)
Output/PromptsINPUTS/COMMENTS


                    <<< LVshim Analysis >>>
(Magnet Type : Oxford (12 coil))

1. Cal File Type (Gradient, Oxf-12 )    : Gradient
2. Generation (Create, Resume or Modify): Create
3. Operation Mode (Service or Research) : Service
4. Calibration File Name Token          : NEW
5. Sampling Diameter in Centimeters     : 45
6. Display Phase Diff Images            : No
7. Display Calibration Matrix           : Yes
8. Control Gradient Shim Coil Current   : Yes

0. Accept (q or s to quit).

Enter the index number to change default (0..9) [0]: .. 1 [Enter] to change Cal File Type
                                                           to Resistive

<<< LVshim Calibration File Generation >>>
    (Magnet Type : Oxford (12 coil))
1. Cal File Type (Gradient, Oxf-12 ) : Oxf-12
2. Generation (Create, Resume or Modify): Create
3. Operation Mode (Service or Research) : Service
4. Calibration File Name Token          : NEW
5. Sampling Diameter in Centimeters     : 45
6. Display Phase Diff Images            : No
7. Display Calibration Matrix           : Yes
8. Show Suggested Perturb Current       : Yes
9. Suggested Perturb Current.

     Z1: 0.1       X: 0.1       XY: 1.0
     Z2: 0.5       Y: 0.1    X2-Y2: 1.0
     Z3: 2.0      ZX: 1.0      Z2X: 2.0
     Z4: 2.0      ZY: 1.0      Z2Y: 2.0

0. Accept (q or s to quit).

Enter the index number to change default (0..10)[0]: ..    [Enter]
Displaying Suggested Perturb Current
                  

Data Collection

Control Gradient Shim Coil Currents

Note:

Option availability I is available only when the Cal File Type is "Gradient."

When this option is set to yes, the gradient shim coil perturb currents are automatically set by the software. When the option is set to no, the gradient shim coil perturb currents must be manually set.

To manually set the gradient shim coil currents:

  1. At the operator workspace, select Manual Prescan.

  2. Set x, y, and z to the desired values, then select Files, then Save Shim File. Select Done to save the file.

  3. Select Done to exit Manual Prescan.

Suggested Perturb Current

Note:

Option availability II is available only when the Cal File Type is not "Gradient."

When this option is set to yes, the suggested perturb currents for the coil set being calibrated are displayed, allowing you to get a hard copy of the currents. This option allows you to change the perturb currents to be used for the coil set being calibrated. When selected, the software prompts for the new currents, one at a time. Illustration 5-2A shows a typical suggested perturb currents display (shown in Genesis format; the Lx display is different).

Figure 2. SUGGESTED PERTURB CURRENTS (OXFORD RESISTIVE SHIM COIL)

Storage

There are two types of storage: Accumulate and Replace. Accumulate is used when you wish to average multiple scans for a given coil type/perturbation (such as taking three Z1 baseline scans to average the data). Replace is used when you wish to throw out the old data and replace them with data from a different scan. If you choose Replace mode, all previous data collected for the coil are lost.

Data Collection

  1. Perform a Phantom Setup, if you haven’t already done so.

  2. At the operator workspace, select the Scan icon in the desktop control panel and set up LVshim scan as follows:

    1. In RX Manager, click New Series.

    2. At the operator workspace, prepare the system for LVshim scanusing the "Service Protocols" procedure located on the service methods CD-ROM, or, for the alternate proprietary procedure, see below.

    3. In Additional Parameters, click User CVs Screen.

    4. In the User Control Variables window enter the following:

      • No. of Scan Planes: 6

      • Bandwidth: 2000

    5. Click Accept and then Save Series.

    6. In RX Manager, click Prepare to Scan.

  3. Click Manual Prescan. While in the Center Freq Coarse (CFC) mode, verify the frequency peak is centered in the display. Adjust the center frequency peak as necessary. After locating the best signal, open the [Frequency] menu at the top of Manual Prescan window and select Save Frequency to save the new center frequency.

  4. Click Scan TR (R1/R2). Verify that R1 and R2 are not over ranged. Adjust R1 and R2 for approximately 50%.

  5. Click Done to exit manual prescan.

  6. Click Scan.

    Note:

    A Resistive Shim Calibration file can be created by using one baseline scan and one perturbation scan for each coil in a coil set. However, if time allows, three scans should be taken for each baseline and perturbation to insure accuracy of the calibration matrix (by averaging the data).

  7. If more than one scan is desired, when the first scan is finished, press Scan again to collect a second set of scan data. Then, when the second scan is finished, press Scan again to collect a third set of scan data.

  8. When the baseline scan(s) is finished, in RX Manager, copy and paste the present series, then select View Edit. Change the Series Description to Z1 0.1, then click Save Series, followed by Prepare to Scan. Perturb the Z1 coil baseline current by 0.1 amps as described in Suggested Perturb Currents and Series Descriptions, then repeat steps 3 through 8 to collect the Z1 perturbation scan(s). After the Z1 perturbation scan(s) have been collected, set the Z1 coil current back to its baseline value.

  9. To collect scan data for all of the coils in a coil set before processing, follow these steps:

    1. In RX Manager, copy and paste the present series, then select View Edit.

    2. Change the Series Description to Z2 baseline, then click Save Series, followed by Prepare to Scan.

    3. Perform steps 3 through 9 to collect baseline and perturbation scans for the Z2 coil (use Z2 0.5 as the Series Description name for the Z2 perturbation scans).

    4. Repeat steps a through c for the remaining coils in the coil set.

    5. When scan data for all coils have been collected, perform Shim Calibration Processing (Silent Mode = Yes).

  10. To collect and process scan data for each coil, one at a time

    1. Perform Shim Calibration Processing on the Z1 coil scans (baseline and perturbation) (Silent Mode = No).

    2. When the processing is done, in RX Manager, copy and paste the present series, then select View Edit. Change the Series Description to Z2 baseline, then click Save Series, followed by Prepare to Scan.

    3. Perform steps 3 through 9 to collect baseline and perturbation scans for the Z2 coil.

    4. Repeat steps a through c for the remaining coils in the coil set.

Existing Current in Each Coil

Existing current values are displayed for all the coils in the coil set being shimmed. This option allows you to change the existing values. When the gradient current values are changed, the new values are automatically sent to the gradient amplifiers by the software.

Shim Calibration Processing

  1. In the Service Browser, select the Calibration tab, then in the menu options on the left side of the browser, click LVShim. On the right side, click the link Click here to start this tool. See Figure 1.

  2. For TwinSpeed, highlight the GradMode and click on OK. Continue as shown in Table 6.

Table 6. SHIM CALIBRATION PROCESSING
OUTPUT/PROMPTSINPUTS/COMMENTS


                      <<< LVshim Calibration File Generation >>>
(Magnet Type : Oxford (12 coil))

1. Cal File Type (Gradient, Oxf-12 )    : Gradient
2. Generation (Create, Resume or Modify): Create
3. Operation Mode (Service or Research) : Service
4. Calibration File Name Token          : NEW
5. Sampling Diameter in Centimeters     : 45
6. Display Phase Diff Images            : No
7. Display Calibration Matrix           : Yes
8. Control Gradient Shim Coil Current   : Yes

0. Accept (q or s to quit).

Enter the index number to change default (0..9) [0] :
                    

1 Enter to change Cal File Type to Oxf-12


                      <<< LVshim Calibration File Generation >>>
(Magnet Type : Oxford (12 coil))
1. Cal File Type (Gradient, Oxf-12 )    : Oxf-12:  
(Additional menu selections displayed)
9. Suggested Perturb Current.

Z1: 0.1       X: 0.1       XY: 1.0        
Z2: 0.5       Y: 0.1    X2-Y2: 1.0        
Z3: 2.0      ZX: 1.0      Z2X: 2.0        
Z4: 2.0      ZY: 1.0      Z2Y: 2.0

  0. Accept (q or s to quit).
Enter the index number to change default (0..10) [0]:
                    

0 Enter when all menu items are acceptable.


                      <<< Characterizing Shimming Coils >>>
  
Z1:         X:          XY:        
Z2:         Y:       X2-Y2:        
Z3:        ZX:         Z2X:        
Z4:        ZY:         Z2Y:      
Common Baseline: Pending
1. Image Data (Ex, Se, Im Number) : 50008, 1, 1  
(Series Description: Z1 baseline)
2. Storage (Accumulate or Replace): Accumulate
3. Display Phase Diff Images      : No
4. Processing in Silent Mode      : No
5. Coil Name (To be Characterized): [None Selected]

0. Accept (q or s to Return).

Enter the Index Number to Change Default (0..6) [0]: ...
                    

5 Enter (To select coil to be characterized.)


                      1.   Z1   0.100 [Z1]  
2.   Z2   0.500 [Z2]  
3.   Z3   2.000 [Z3]  
4.   Z4   2.000 [Z4]  
5.    X   0.100 [X]  
6.    Y   0.100 [Y]  
7.   ZX   1.000 [ZX]  
8.   ZY   1.000 [ZY]  
9.   XY   1.000 [XY]
10. X2-Y2  1.000 [X2_Y2]
11.   Z2X  2.000 [Z2X]
12.   Z2Y  2.000 [Z2Y]  
0.  Common Baseline.

Enter the Index Number (0..12) [6] : ..................
                    

1 Enter to select Z1 coil


                      <<< Characterizing Shimming Coils >>>
  
Z1:         X:          XY:        
Z2:         Y:       X2-Y2:        
Z3:        ZX:         Z2X:        
Z4:        ZY:         Z2Y:      
                    


                      Common Baseline: Pending
1. Image Data (Ex, Se, Im Number):50008, 1, 1 <-------
                    

Be sure this is the correct (Series Description: Z1 baseline) image data set for the Z1 baseline scan


                      (Series Description: Z1 baseline)
2. Storage (Accumulate or Replace): Accumulate
3. Display Phase Diff Images     : N
5. Coil Name (To be Characterized): Z1
                    

6.Perturbation Current : 0.100 <--------------- 0 Enter to change Perturbation Current to Baseline
0. Accept (q or s to Return).
Enter the Index Number to Change Default (0..7)[0]: .. 0 Enter when all of the menu items are acceptable.


                      If Silent Mode was set to Yes, the following is displayed:
Processing ... plane  1 of 6.
                    

Image data are processed (planes 1 - 6)
Processing ... done. 6 of 6
When all the coils have been characterized, the following is displayed (if Yes was entered for Display Calibration Matrix):
Displaying Calibration Matrix
Cal File Generation Complete. Run Again? (Y,N)[N]: ...Enter y or n as appropriate
If Silent Mode was set to No, the following is displayed:
Processing plane 1 of 6 <---------------------Image data is processed (planes 1 - 6).
Processing ... done. 6 of 6
If image phase wrap is detected, the following message appears:


                      Excessive phase wrapping
....Please re-scan at Wider Bandwidth or smaller test diameter
                    

Exit Cal File Generator? (Y,N) [N]: ..................Select Research Operations and Modify CVs. Type opuser2 and enter a higher bandwidth. Valid bandwidth entries are: 200, 500, 1000, and 5000. Select Research OperationsDownload, and restart the calibration file process using the higher bandwidth.
If Yes was entered for Display Phase Diff Images, the following appears:


                      The Following Are the Image Display Options.
1. Image Plane 1
2. Image Plane 2
3. Image Plane 3
4. Image Plane 4
5. Image Plane 5
6. Image Plane 6
7. Image With Sampling Pixels.
8. Save Phase Difference Image Into Database.
0. Return
                    

Enter Index (0..8) [1] : .............................Enter appropriate choice for more option information, or 0 Enter to continue.


                      Coil Name (To be Characterized)  : Z1 <---------------
Perturbation Current             : 0.100
                    

When processing is done, the next image data set is displayed and processed. Each image data set is displayed and processed until all have been processed. If all the data sets have been processed, but not all the coils have not been characterized, the Characterizing Shimming Coils screen is displayed.
More Coils to Process (Y,N) [Y] : .................... y Enter to process the Z1 perturbation scan (or a second Z1 baseline scan).
If Yes was entered at More Coils to Process prompt, the following appears:


                      <<< Characterizing Shimming Coils >>>
Z1: (B 1, C 0) X: XY: <---------
Z2:    Y:    X2-Y2:
Z3:    ZX:    Z2X:
Z4:    ZY:    Z2Y:
Common Baseline: Pending
                    

The software automatically shows the current status of the coil characterization process. In this example, B 1 means that one baseline scan has been performed, and C 0 means that zero perturbation scans have been performed.


                      1. Image Data (Ex, Se, Im Number): 50008, 2, 1 <-----  
(Series Description: Z1 0.100)
2. Storage (Accumulate or Replace): Accumulate
3. Display Phase Diff Images      : No
4. Processing in Silent Mode      : No

                    

The software automatically increments to the next image data set. If there are no more LVshim images, it displays [None Selected]. Be sure that these are the correct data before accepting.


                      5. Coil Name (To be Characterized): Z1  <----
6. Perturbation Current           : 0.100  <-------

0. Accept (q or s to Return).
                    

The software automatically increments to the next logical coil name and perturbation current. Be sure this information is correct before accepting
Enter the Index Number to Change Default (0..7)[0]: .. 0 Enter when all menu items are acceptable.
Processing plane 1 of 6 <---------------------Image data are processed (planes 1 - 6).
Processing ... done. 6 of 6
More Coils to Process (Y,N) [Y] : ....................Repeat scanning and processing until all coils in the coil set have been characterized, then nEnter to continue.


                      When No is entered at the More Coils to Process prompt, this happens:
If some coils have not been characterized, the following appears:
                    


                      Coil Characterization Is Incomplete.
Not Enough Data To Create Calibration Matrix
Build Partial Matrix? (Y,N) [Y] : ....................
                    

Since not all of the coils have been characterized, the software may not able to create a calibration file. Generally, you should enter n at this prompt.
If No is entered at Build Partial Matrix prompt, the following appears:


                      Process Will Be Saved and Can Be Resumed Later.
Exit Cal File Generator? (Y,N) [N] : .................
                    

Enter y or n as appropriate. The processed calibration data are saved. The calibration generation process can be resumed later by selecting Resume for menu item 2 (Generation).


                      Coil Characterization Is Incomplete.
Matrix Is Not Invertable.
Process Will Be Saved and Can Be Resumed Later.
Exit Cal File Generator? (Y,N) [N]: ..................
                    

Enter y or n as appropriate. The software was not able to generate a calibration matrix because some of the coils have not been characterized. The processed calibration data are saved. The calibration generation process can be resumed later by selecting Resume for menu item 2 (Generation).


                      If all the coils have been characterized, the following appears
(if Yes was entered for Display Calibration Matrix):
                    


                      Displaying Calibration Matrix
Displaying Calibration Matrix
Please Get the Hardcopy if Desired.
Press Enter to Continue. [ ] : .....................
                    

The Calibration Matrix is displayed. See Section 6-9 for details. If a hard copy of the matrix is desired, print it now. Then, press Enter to continue.
Cal File Generation Complete. Run Again? (Y,N) [N]: ..Enter y or n as appropriate.
Note:

Signa 1.0T Mobiles may have a Z2 Resistive Shim Coil installed on the Passive Shim Drum. Make sure that the Z2 Resistive Shim Power Supply (mounted on the operator workspace table) is powered off when performing this procedure.

Sampling Diameter in Centimeters

The sampling diameter is the size of the diameter spherical volume (DSV) to be used during processing. Only data within this specified volume are processed. This value is editable during creation of the shim calibration file, but is fixed during normal shimming.

Details of Characterizing Shimming Coils

This section provides detailed information about the items in the Characterizing Shimming Coils menu. When selected, some of the items toggle to another value (available choices are shown in parentheses), some toggle between Yes and No, and some prompt for additional input.

Coil Characterization Status

The coil characterization status is located in the area under the menu title line. It consists of the following information for each of the coils in the coil set:

  • Coil name

  • Number of Baseline scans for the coil (designated with a B)

  • Number of Perturbation scans for the coil (designated with a C)

The following is an example of the coil characterization status for a typical coil (with three baseline scans and three perturbation scans):

AX_1: ( B 3, C 3 )

The coil characterization status area also shows whether a common baseline has been processed for the coil set. A common baseline is a baseline scan used during characterization of all the coils in a coil set. A common baseline is not necessary if a new baseline scan is performed before each coil perturbation.

Coil Name

This is the name of the coil to be characterized by the selected image data set. If the selected image data set is being used as the baseline scan for all the coils, use the name common baseline.