• Topic ID: id_2001435
  • Version: 7.0
  • Date: Dec 5, 2019 2:31:41 AM

Disabling TR and RF inputs for body mode

Procedure

  1. Make sure that the system is idle and all coils have been removed from the magnet bore.
  2. Open the front door of the .
  3. Disable the following switches on the driver module and the exciter.
    1. On the front of the driver module, set SW2 to TR DISABLE (down).
      note: Leave SW1 set to HV ENABLE (up).
    2. On the front of the exciter, set the RF ENB switch to the disabled (down) position.