• Object ID: 00000018WIA301F0130GYZ
  • Topic ID: id_12373879 Version: 1.11
  • Date: Aug 26, 2020 1:16:39 PM

Scan Room Blower Box Replacement

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
2Not Applicable30 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Non-magnetic tool set1---
Table 3. Consumables
ItemQuantityEffectivityPart numberManufacturer
Cable ties (standard 4-6 inch)----
Electrical insulating tape----
Table 4. Replacement parts
ItemQuantityEffectivityPart numberManufacturer
Dual blower box1RRx receive chain systems

5213397–100

-
Table 5. Safety
Table 6. Required conditions
ConditionReferenceEffectivity

All persons performing this procedure must complete GE-certified MR safety training.

--

Power Down and LOTO

Procedure

Perform LOTO on the PEN cabinet. See the MR Service Safety Manual, PN 5452735.

Removing Blower Box

Procedure

  1. Allow five minutes for the filters to discharge. After five minutes, check capacitors for discharge with grounded cable on DVM.
  2. Remove any cable ties on the blower box power cable.
  3. In the magnet room, remove the two top screws and two bottom screws from the blower box power supply cover on the lower right corner of the PEN panel.
    Figure 1. Blower Box Power Supply Cover
  4. Be sure the power cables are properly marked or labeled for later reconnection. Disconnect the power cable connectors from the AC power filters. See Table 7for a list of cables and connections.
    Note:

    The single blower box has five power cables. The dual blower box has seven power cables.

    Figure 2. AC Power Filters
  5. Follow the cable path and confirm that the cable is free of the penetration wall and will not tangle when the blower box is removed.
  6. Remove both air hoses at the top of the blower box.

    Systems with the RRx receive chain use the dual blower box, which uses two hoses to provide air for both patient comfort and the RF hub.

    Note:

    Mark or label the two air hoses at the top of the blower box to note their mounting location.

    Figure 3. Air Hose Clamps
  7. Note:

    Some sites have a closet service hatch installed in the scan room wall. Use the closet service hatch to safely remove the blower box from the scan room.

    Loosen, but do not remove, the four screws securing the blower box to the PEN panel.
    Note:

    If the slot on the mounting flange does not provide enough clearance for placing over the existing flange studs, remove the mounting screws, position the blower box on the SPW, and attach with the removed screws.

    At least two people are required for this procedure.

    Figure 4. Side View of Blower Box
  8. With two people holding the blower box, exit the magnet room by walking as close to the wall on the coldhead side of the magnet as possible. See Figure 5.

Installing New Blower Box

About this task

The following steps explain how to install the scan room blower box.

Procedure

  1. With two people holding the blower box, stay as far away from the magnet as possible as you bring the blower box into the magnet room and position it on the PEN panel. See Figure 5. Refer to note inStep 8 above to determine safe distance.
    Note:

    Some sites have a closet service hatch installed in the scan room wall. Use the closet service hatch to safely insert the blower box into the scan room.

  2. With two people, lift and slide the blower box mounting holes over loosened screws on the PEN panel
    Note:

    If you had to remove one or more of the screws during the removal process, one person must hold the blower box in place while the second person installs the screws.

  3. With one person holding the box in place, tighten the four screws to secure the blower box to the PEN panel frame.
    Figure 5. Mounting New Blower Box
  4. Attach the blower box power cable to the power supply panel filters, matching the labels on each connector to its corresponding filter.
    Important: Arc flash may occur if exposed terminals contact AC power filter connections.
    Note:

    There should be a washer above and below each ring terminal on the filter stud.

    Figure 6. AC Power Filter Cable Connections
    Table 7. Cables and Connections

    Power cable color

    PEN panel AC power connection

    Blower type (single, dual, or both)

    Blue

    PEN-F11

    Both

    Yellow

    PEN-F10

    Both

    Red

    PEN-F9

    Both

    White

    PEN-F8

    Both

    Green and yellow

    PEN-F7

    Both

    Brown

    PEN-F6

    Dual blower only

    Black

    PEN-F5

    Dual blower only

  5. Replace the power supply cover. Ensure the cable is routed through the cable access hole, and is not pinched under the cover. The warning label should be upright.
    Note:

    The length of the power cable from the bottom of the blower box to the power supply filter panel provides enough slack to run the cable along the right side of the PEN panel using cable ties to hold the cable in place.

  6. Replace the air hose connections, noting the correct location. Secure the air hose by tightening the worm gear hose clamps. Pull on each hose to confirm the hose is securely fastened to the blower box.

Finalization

Procedure

  1. Remove LOTO from the PEN cabinet. See the MR Service Safety Manual, PN 5452735.
  2. Ensure air is flowing to the patient bore.
  3. (For systems equipped with the legacy RF hub) : Ensure air is flowing to the RF hub.