- 00000018WIA3062C630GYZ
- id_20116294.1
- Sep 2, 2020 3:24:33 PM
Replacing and cleaning the blower box filter
Clean and replace the blower box filter.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 20 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Nonmagnetic Tool Kit, Inch/Metric | 1 | 46-320273G4 | - |
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Charcoal Media Filter | AR | 5265337 | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
Procedure
Visually inspect the blower box filter(s) for dust. If a filter requires cleaning or replacing, do the procedures that follow in this document.