• Object ID: 00000018WIA307C3930GYZ
  • Topic ID: id_2012445 Version: 4.1
  • Date: Sep 23, 2019 3:48:38 PM

Adjusting the static pressure on the compressor

When both the static and dynamic compressor pressures are out of range, adjust the static pressure.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes -

About this task

When both the dynamic and static compressor pressures are out of range, adjust the static pressure to conform with the Gas Pressure Adjustment instructions in the appropriate publication below.
Manual titleManual number
Operation Manual SRDK Series Cryocooler (For F-50 Series Compressor Unit)CD32ZZ-271
Operation Manual SRDK Series Cryocooler (For CSA-71 Series Compressor Unit)CD32ZZ-056
Magnet and Cryogen Manual for Passively Shimmed Magnets5495018
Magnet and Cryogen Manual for Actively Shimmed Magnets5495019
Technical Instruction RDK-408A2 4K ColdheadCD32ZZ-098
Technical Instruction RDK-408A3 4K ColdheadCD32ZZ-372
Technical Instruction RDK-408T 10K ColdheadCD32ZZ-058
Technical Instruction CSA-71A Compressor UnitCD32ZZ-60