Collimator Assembly Replacement
Prerequisites
Overview
![]() |
|
note: Please check the component P/N when replace the collimator related parts, should be ordered the same P/N according to the label on each component.
Procedure
- Move table to its lowest elevation.
- Remove gantry covers as required.
Refer to
- Position the tube at 3 o’clock.
danger- Turn OFF all three (3) gantry service switches (Axial Drive Enable, HVDC Enable, 120 VAC). Perform Lockout/Tagout procedures to remove all power from gantry.
danger- Engage gantry rotational lock.
warning- Connect and secure hoist and tube change bracket to the tube.
- Remove high voltage cables from cathode and anode using spanner wrench.
- Remove the tube by following the procedure in Tube Replacement.
- notice
- Disconnect connectors J2, J3 and J11 from collimator.
- Remove collimator by removing the two (2) mounting bolts from the top, two (2) bolts from the bottom, and 2 bolts from the rear of the collimator.
- Install new collimator by replacing the six (6) bolts removed
in above step.
- Apply Pre-load torque to M14 Mounting bolts:
- Apply Final torque to M14 Mounting bolts:
- Apply Pre-load torque to M14 Mounting bolts:
- Secure the tube back in place.
- Replace high voltage cables back using spanner wrench.note: If oil needs to be topped off, be careful of spills. Do not use any part of the gantry as a shelf to rest oil on.
- Reassemble gantry.
![]() |
![]() |
Finalization
- Perform Flash Download.
- Perform Gantry Rotation Safety Check.
- Perform Gantry Balance Procedure.
- Perform Collimator Aperture Test.
- Perform Collimator Functional Test.
- Perform Plane of Rotation (POR) Alignment.
- Perform Beam on Window (BOW) Alignment.
- Perform CBF/SAG Alignment Process.
- Perform ISO Alignment Procedure.
- Perform Hot ISO Alignment.
- Perform HHS Scans
- Perform Collimator Calibration.
- Perform Detailed Calibration.
- Perform Fast Calibration.
- Perform System Scanning Test.
- Perform Quality Assurance Test.
- Perform Save System State.