- Topic ID: id_11038689
- Version: 3.0
- Date: Sep 13, 2021 2:19:15 AM
Quality Assurance Test
Prerequisites
This document provides the necessary steps to ensure system image quality before releasing.
1 Phantom Setup
Procedure
- Record the phantom serial number on GE e4879 Component Replacement Form.
- Perform Center Phantom
- Set Internal Landmark.
2 Acquire Scans
Procedure
- Select NEW PATIENT from bottom of Scan monitor.
- Type the following entries in the listed Patient Information fields:
- Patient ID: Service
- Patient Name: Quality Assurance Test
- Select Anatomical Selector Service tab from Protocol Selection.
- Select MANUFACTURING, IMGSER 20CMQA from Anatomical Selector Service tab.
- Select CONFIRM.
- Press START on SCIM / GSCB when it flashes. Scans are performed.
- Select NEXT SERIES.
- Select CONFIRM.
- Press MOVE TO SCAN on SCIM /GSCB if it flashes.
- Press START on SCIM / GSCB when it flashes. Scans are performed.
- Repeat Steps [7] – [10] until all series have been performed and completed.
- When all series are completed, select END EXAM.
- If a PPS Response window is displayed, select COMPLETE.
3 Analysis
Procedure
- Select SERVICE desktop and move SERVICE desktop window to left monitor.
- Select IMAGE QUALITY tab from SERVICE desktop.
- Select IMAGE ANALYSIS and move IMAGE ANALYSIS window to left monitor.
- Select scanned examination from right monitor List/Select browser.note:
Series 1 and Image 1 should be selected. If not, select them.
- Press AUTO 1X radio button on IMAGE ANALYSIS window.
- Select AUTO 1X selection button, IMGSER 20QA, and first test choice.note:
Number of test choices correspond to the number of series in the scanned examination. The order of the test choice correlates with the series number of the scanned examination (Series 1/Test 1, Series 2/Test 2, etc.).
- Select ACCEPT.note:
At test completion, test data results, including specification limits and Pass/Fail results, are displayed.
note:At the top of the results window, the test Pass/Fail summary result is posted.
- Record results on GE e4879 Component Replacement Form.
- Click DISMISS from Report window.
- Select next series from right monitor List/Select browser.
- From IMAGE ANALYSIS window, select AUTO 1X selection button, and select IMGSER
20QA and next test choice.note:
If test choice has multiple selections, execute each selection.
- Select ACCEPT.note:
If test requires user input, follow instructions in IMAGE ANALYSIS window lower right pane.
note:At test completion, test data results, including specification limits and Pass/Fail results, are displayed.
note:At the top of the results window, the test Pass/Fail summary result is posted.
- Record results on the GE e4879 Component Replacement Form.
- Click DISMISS from Report window.
- Repeat steps [10] – [14] until all tests have been performed.
- When all tests are completed, click QUIT.note:
If any test fails, repeat failed test. If the test fails persistently, perform appropriate corrective action.
4 Image Review for Bands and Streaks
Procedure
- From IMAGEWORKS desktop List/Select browser, select scanned examination, and Series 3.
- Select VIEWER and review all images from series for band and streak artifacts.
- If any images depict bands, perform steps in Band Analysis, Band Analysis, on those images.
- If any images depict streaks, perform Streak Test.
- In viewer, select series 5 and review all images from series for band and streak artifacts.
- If any images depict bands or streaks, repeat Step [3] or [4].
- Repeat Steps [5] – [6] for Series 7 and 10.
5 Band Analysis
Procedure
- Select SERVICE desktop and move SERVICE desktop window to left monitor.
- Select IMAGE QUALITY tab from SERVICE desktop.
- Select IMAGE ANALYSIS and move IMAGE ANALYSIS window to left monitor.
- From the List/Select browser on right monitor, select image to be analyzed.
- Press MANUAL radio button on the IMAGE ANALYSIS window.
- Click MANUAL selection button, then BAND.
- Click ACCEPT. The image is displayed in IMAGE ANALYSIS window with a Band ROI.
- Follow instructions, in IMAGE ANALYSIS window lower right pane, to size Band ROI and analyze.
- From results, if Avg. Ref. Band Factor is out of specification range of +/-2.8, perform appropriate corrective action.
- When all tests are completed, click QUIT.