• Topic ID: id_18717851
  • Version: 1.0
  • Date: Sep 20, 2018 2:04:10 PM

PM Schedule - BrightSpeed Elite, Edge, Excel

These schedules are for BrightSpeed only, as approved by CT Engineering. BrightSpeed PMs are based on gantries doing 2 million revolutions per year.

NOTE: last Revised: 28 Oct. 2015

NOTE: The PM is based on (3) service cycles per year at 670k revolution intervals.

  • Schedule A - PM cycle begins when Rev Counter shows xxx670k revs.

  • Schedule B - When Rev Counter increments another xxx670k revs.

  • Schedule C - When Rev Counter increments another xxx670k revs.

NOTE: PM time without camera (includes time allocation for Gantry cover handling):

  • Schedule A - 4.0 hours

  • Schedule B - 3.0 hours

  • Schedule C - 4.0 hours

NOTE: ** This step applies only to systems with filters on the DAS duct fan assembly.

Console

Gantry

High Voltage

PDU

System

Table