• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA30E18650GYZ
  • Topic ID: id_2024769 Version: 2.1
  • Date: Jun 1, 2020 3:15:36 PM

Installing the exciter

Install a replacement exciter.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-

Procedure

  1. Install the new exciter in the cabinet.
  2. Connect all the cables to the front and back of the exciter as they were prior to removing.
  3. Tighten the thumbscrews to secure the exciter to the PEN cabinet.
    Note: After LOTO is removed and power is reapplied to PEN cabinet, allow 30 minutes for the exciter to reach operating temperature and stabilize before calibrating or scanning.