- SIGNA MR355 / SIGNA MR360
- Service Manual
- 5856356-3EN Revision 5.0
- Basic Service Documentation. Copyright General Electric Company.
- 00000018WIA30388F20GYZ
- id_131067622.0
- Jul 19, 2019 11:00:42 AM
Universal Power Monitor - Body Setup and Calibration
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 30 minutes | Not Applicable |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| 50 Ohm Terminator | 1 | - |
46-265874p1 | - |
| 7/16Male -N Female Adapter (Attached in System Cabinet) | 1 | - |
5166139 | - |
| ||||
About this task
This procedure sets up the Universal Power Monitors (UPM’s) in Body Mode, by using the system and the Power Measurement Kit (Wattmeter). For proper calibration of the Universal Power Monitors, (UPM) Forward Power must be completed in Body mode.
BODY FORWARD POWER CALIBRATION
Procedure
Setting up Software Tool to Calibrate UPM - Body Forward Power
About this task
This tool must see a Landmark only. It is not necessary to have a Head coil in place, or any other scan parameter entered than what is described below. Simply Landmark on the cradle where the Head coil would be.
Procedure
If Head UPM is NOT calibrated before Body UPM Calibration, reset “UPM1Cal.cfg” and “UPM2Cal.cfg” file to 0 according to the following procedure.Notice - Note:Start a New Exam
This tool must see a Landmark only. It is not necessary to have a Head coil in place, or any other scan parameter entered than what is described below. Simply Landmark on the cradle where the Head coil would be.
- Enter the following parameters:
- geservice
- Weight: 111
- Press the Align button on, (Land Mark).
- Move the cradle to the Home position and then drive it into the bore.
- Press Landmark button when the align light is approximately where the center of the head coil would be.
- From the Common Service Desktop Select:
- Select the Nuclei 1H.
- Select the RF Chain you are going to calibrate Body.
- Select Power Monitor type. Forward
- Select: Calibrate Button
- Select the Start button on the tool opened earlier, Figure 13.
- Status should indicate Pass/Fail from 1 minute to 7.5 minutes depending on software version (tool should notify user of current step in progress). If failure occurs, see Universal Power Monitor (UPM) Troubleshooting.
Finalization
Procedure
- Remove the cables and power measurement equipment.
- Re-enable T/R Drive . See Driver Module, Figure 1.
- Upon completion of UPM calibration, Reset TPS. The system will not scan until this step is done











