• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30A36E20GYZ
  • Topic ID: id_13107612 Version: 3.0
  • Date: Aug 29, 2019 1:35:13 AM

Table Top Height Adjustment

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1 - - -

Procedure

  1. Dock Patient Transport.
  2. Pump table top to fully raised position.
  3. Use a small screwdriver and carefully remove cover plug on table top. See Figure 1.
    Figure 1. REMOVING COVER PLUG
  4. Adjust screw in table top to set table top height equal to bridge height.
  5. Replace cover plug.

Finalization

No finalization steps.