PROCEDURE TO TAKE TABLE OUT FROM THE MAGNET ROOM
Prerequisites
Table 1. Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
2
0 minutes
30 minutes
0 minutes
Table 2. Tools and test equipment
Item
Quantity
Effectivity
Part number
Manufacturer
Standard Tool (non-ferrous)
1
-
-
-
Table 3. Safety
Warning
DO NOT TAKE ANY FERROUS TOOLS INSIDE THE MAGNET ROOM FORCE
OF THE MAGNETIC FIELD OF THE MR SYSTEM CAN CAUSE FERROUS OBJECTS TO
BECOME PROJECTILES THAT CAN CAUSE SERIOUS INJURY.
Warning
AT ANY TIME, DO NOT TAKE THE TABLE BEYOND THE MARKED LINE
AS SHOWN IN
THE FORCE OF THE MAGNETIC FIELD OF THE MR SYSTEM CAN CAUSE
TABLE ACTUATOR NOT TO FUNCTION.
While servicing the table it may be required to disconnect
the table from Gantry and required to take out from the magnet room.
This instruction provides procedure to take out the table out from
the Magnet Room
Procedure
Remove FRP bottom covers of Fixed Table.
Remove the Rubber cap and then screws of the FRP bottom cover
on the near the Gantry end of the Table.
Figure 1. FRP Cover1
Similarly remove the cap and screws of the FRP bottom cover
on the rear end of the Table.
Figure 2. FRP Cover2
Remove the right Bottom FRP Cover. Similarly remove the left
Bottom FRP cover on the other side
Figure 3. FRP Cover3
Now remove the Dock FRP Cover Locking screws and then the Dock
FRP cover carefully.
Note: There are cable below the FRP cover. Be careful not to damage them.
Figure 4. FRP Cover4
Keep both Bottom FRP Covers and Dock FRP covers aside safely.
Disconnect cables.
Disconnect the Sensor cable and Actuator Power supply cable
from the Power supply extension cable from Control box.
Figure 5. Disconnect cable 1
Disconnect the Earthing cables from the Dock Frame.
Figure 6. Disconnect cable 2
Disconnect the Control Board Cable-2 (TEB Cable) Assembly from
the connector box.
Figure 7. Disconnect cable 3
Disconnect the Control Board Cable (TEB Cable) Assembly from
the connector box.
Figure 8. Disconnect cable 4
Remove the screws and nuts connecting the Table and the Dock
frame.
Figure 9. Dock Frame
Mark the position of the 4 pads so that it can be restored
later.