Cabinet Power-Up Procedure

Prerequisites

About this task

Procedure

Before power is reapplied to the Systems Cabinet, please verify the following conditions exist:
  • All circuit boards are seated properly into the backplane, and the faceplate screws are tightened.
  • All cable connections are in place and secured properly.
  • All covers and panels have been secured properly, and there is no loose hardware resting on any of the surfaces.
  • Restore power and remove LOTO on the RF Amp/Systems Cabinet/Pen Panel. See the MR Service Safety Manual, PN 5452735.

Finalization