o.19.2 LVShim Scan
SCAN OPTION INPUT
PATIENT REGISTER . New Pt
Patient ID geservice
Weight (Lb) 300
Landmark
Landmark >Sternal
Patient name LV Shim
PATIENT POSITION
Patient Position >Supine
Patient Entry >Head First
Coil ...BodyAccept
Series Description Note 1
IMAGING PARAMETERS
Plane >Axial
Mode >2D
Pulse Seq ...Spin EchoAccept
For Twin Speed ONLY: Grad Mode Whole
Imaging Options none (default)
Psd Name LVshim
Protocol no entry
ADDITIONAL PARAMETERS USER CVs
USER CONTROL VARIABLES
Bandwidth in Hz Note 2; 100.0 to 20000.0
Accept
(lowest window) Save Series
Note:
1: Refer to procedure for description name to use.
Note: