• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30BD6E20GYZ
  • Topic ID: id_13106882 Version: 4.0
  • Date: Mar 23, 2020 2:36:44 PM

o.19.2 LVShim Scan

SCAN OPTION INPUT

PATIENT REGISTER . New Pt

Patient ID geservice

Weight (Lb) 300

Landmark

Landmark >Sternal

Patient name LV Shim

PATIENT POSITION

Patient Position >Supine

Patient Entry >Head First

Coil ...BodyAccept

Series Description Note 1

IMAGING PARAMETERS

Plane >Axial

Mode >2D

Pulse Seq ...Spin EchoAccept

For Twin Speed ONLY: Grad Mode Whole

Imaging Options none (default)

Psd Name LVshim

Protocol no entry

ADDITIONAL PARAMETERS USER CVs

USER CONTROL VARIABLES

Bandwidth in Hz Note 2; 100.0 to 20000.0

Accept

(lowest window) Save Series

Note:

1: Refer to procedure for description name to use.

Note: