• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30AEBF20GYZ
  • Topic ID: id_13107223 Version: 1.1
  • Date: Jul 6, 2019 12:03:28 AM

o.12.4 Head Slice Thickness (2D, Grad Echo, 5mm)

SCAN OPTION INPUT

RX MANAGER Place cursor in Series list area.

Press right mouse button, then

Copy Series, Paste Series.

Double-click left mouse button

on new series to activate it.

See Note 1

IMAGING PARAMETERS

Pulse Seq ...GREAccept

SCAN TIMING

TE 9

TR 100

Flip Angle 30

Bandwidth 15.63

SCANNING RANGE

Slice Thickness 5

Spacing 77

S/I Start 0

S/I End 0

# of Slices 1

ACQUISITION TIMING

Phase 128

NEX 1 (0:16)

(lowest window) Save Series

RX MANAGER Prepare to Scan

Note:

Only values different from previous series shown. All other values should be the same as the previous series.