Slice Offset Checks Data Sheets

Use this data sheet for BRM-based systems, or for TwinSpeed, GradMode=WHOLE.

Axial Scan
Study/Series/imagesTable DeltaSlice OffsetMeasured Offset

(mm)

Acceptance Specifications
____/ ____/ __/__50mm0 to S100 offset (scan #1)mm100 +/- 4mm
____/ ____/ __/__-50mm0 to I100 offset (scan #2)mm100 +/- 4mm
3D Scan
Study/Series/imagesTable DeltaSlice OffsetMeasured Offset

(mm)

Acceptance Specifications
____/ ____/ __/__50mm0 to S100 offset (scan #4)mm99 +/- 4mm
____/ ____/ __/__-50mm0 to I100 offset (scan #5)mm99 +/- 4mm
Note: Analize slices nearest to I50 and S50

For TwinSpeed, GradMode=ZOOM

Axial Scan
Study/Series/imagesTable DeltaSlice OffsetMeasured Offset

(mm)

Acceptance Specifications
____/ ____/ __/__50mm0 to S100 offset (scan #1)mm100 +/- 4mm
____/ ____/ __/__-50mm0 to I100 offset (scan #2)mm100 +/- 4mm
3D Scan
Study/Series/imagesTable DeltaSlice OffsetMeasured Offset

(mm)

Acceptance Specifications
____/ ____/ __/__50mm0 to S100 offset (scan #4)mm99 +/- 4mm
____/ ____/ __/__-50mm0 to I100 offset (scan #5)mm99 +/- 4mm
Note: Analize slices nearest to I50 and S50