- SIGNA MR355 / SIGNA MR360
- Service Manual
- 5856356-3EN Revision 5.0
- Basic Service Documentation. Copyright General Electric Company.
- 00000018WIA30EA8F20GYZ
- id_131075962.0
- Jul 19, 2019 11:00:55 AM
Sample System_Health.log Output Format
> more system_health.log Enter
System Health script run on:
01/22/03 12:39:40
hostname = b5
Hospital Name = b5
--------------------------------------------------
Config File Checks
--------------------------------------------------
System software revision = 10.0_M4_0303.a
Operating System Version = 6.5
Field Strength = 15000
SR Mode = 150
WHOLE Body SR Mode = 77....................PASS
ZOOM Body SR Mode = 150....................PASS
scannerMode = 0..................PASS
fixedTable = n..................PASS
Configured for broadband spectroscopy = NO
Governing Body = fda2
absoluteAvgSar = 2.0
discreteAvgSar = 2.0
absolutePeakSar = 8.0
discretePeakSar = 8.0
dbdtMHW = 20
absoluteDbdtIEC = 66
discreteDbdtIEC = 20
dbdtFDA1 = 45
dbdtFDA2 = 66
Units for Dbdt research is %
absoluteDbdtResearch = 66
discreteDbdtResearch = 66
Magnet type = 1.5T LCC
Magnet Serial Num = R2467
Multi Coil = YES
RF Amp type = 1.5T SRFD2
Grad Coil Type = TRM
GradWarp coeff file correct...........................PASS
Grad Amp Type = ACGD Gradients
Fast TG is enabled in /w/config/prescan_config.cfg
No functions are emulated out in /w/config/mgd_stage
ampCalHead = 0 Expected = 0...........................PASS
ampCalBody = 0 Expected = 0...........................PASS
psdRfWait WHOLE = 52 Expected = 52..................PASS
psdRfWait ZOOM = 52 Expected = 52..................PASS
psdGradWait WHOLE = 68 Expected = 68..................PASS
psdGradWait ZOOM = 68 Expected = 68..................PASS
Grad cal values of 26128 26158 25885 all pos..........PASS
Grad cal values of 25345 25628 25319 all pos..........PASS
Magnet ramp polarity = pos
WHOLE X axis rise times = 288 Expected = 288.............PASS
ZOOM X axis rise times = 268 Expected = 268.............PASS
Maximum Standard Receiver Bandwidth (kHz) = 250.0
RF amp safety limits in MR config file are correct
magnetEnclosureType=1 - Wide Open Enclosure
boreTempLev1=25
boreTempLev2=29
boreTempHyst=3
boreTempOffset=0
**************************************************
--------------------------------------------------
Video Monitor Data
--------------------------------------------------
System is configured with a color monitor and a color colorPalette.
Display type is LCD with LCD model = NEC1850X
**************************************************
--------------------------------------------------
Coil Config File Checks
--------------------------------------------------
This system has the default coil config template files
All the individual coil config fields will be verified
--------------------------------------------------
The Coil named HEAD is a GE default coil name
fastTGstartTA=120 Expected=120......PASS
mcReconEnable=0 Expected=0........PASS
mcPortEnable=0 Expected=0.........PASS
mcErrorEnable=0 Expected=0........PASS
cableLoss=1.3 Expected=1.3......PASS
xmitAtten=0 Expected=0............PASS
startFastRec=4 Expected=4.........PASS
multiCoil=no Expected=no........PASS
linearQuad=1 Expected=1...........PASS
endFastRec=4 Expected=4...........PASS
fastTGstartRG=12 Expected=12......PASS
extremity=no Expected=no..........PASS
numRec=0 Expected=0...............PASS
coilType=1 Expected=1.............PASS
startRec=0 Expected=0.............PASS
coilLoss=0.032 Expected=0.032.......PASS
headDefaultFreqDir=1 Expected=1...PASS
mcChannelSel=0 Expected=0.........PASS
autoshimRcvr=-1 Expected=-1.......PASS
scic_a=0 Expected=0..............PASS
scic_s=0 Expected=0..............PASS
scic_c=0 Expected=0..............PASS
mCoilSeparationDir=0 Expected=0..PASS
maxFOV=480.0 Expected=480.0......PASS
assetThresh=10000 Expected=10000..PASS
endRec=0 Expected=0.............PASS
mcBiasEnable=0 Expected=0.........PASS
mcVoltConfig0=0 Expected=0.........PASS
mcVoltConfig1=0 Expected=0.........PASS
mcOpenCktConfig0=0 Expected=0......PASS
mcOpenCktConfig1=0 Expected=0......PASS
mcOpenCktConfig2=0 Expected=0......PASS
mcOpenCktConfig3=0 Expected=0......PASS
mcShortCktConfig0=0 Expected=0......PASS
mcShortCktConfig1=0 Expected=0......PASS
mcShortCktConfig2=0 Expected=0......PASS
mcShortCktConfig3=0 Expected=0......PASS
mcSwitchSel=0 Expected=0...........PASS
--------------------------------------------------
Skipping all the other listed coils
--------------------------------------------------
EPI Calibration File Checks
--------------------------------------------------
b0vectors.dat.body.WHOLE file is present in /usr/gvx/bin/......PASS
b0vectors.dat.head.WHOLE file is present in /usr/gvx/bin/......PASS
WHOLE Body start esp=960
WHOLE Head start esp=960
b0vectors.dat.body.ZOOM file is present in /usr/gvx/bin/......PASS
b0vectors.dat.head.ZOOM file is present in /usr/gvx/bin/......PASS
ZOOM Body start esp=928
ZOOM Head start esp=928
delay.dat.WHOLE file present in /usr/gvx/bin/..................PASS
X_delay_whole=-4 Y_delay_whole=-2 Z_delay_whole=-5
delay.dat.ZOOM file present in /usr/gvx/bin/..................PASS
X_delay_zoom=-1 Y_delay_zoom=-1 Z_delay_zoom=-4
delay.espb.xyz file present in /usr/gvx/bin/.............PASS
delay.esph.xyz file present in /usr/gvx/bin/.............PASS
delay.espb.xyz.WHOLE file present in /usr/gvx/bin/.......PASS
delay.espb.xyz.ZOOM file present in /usr/gvx/bin/.......PASS
delay.esph.xyz.WHOLE file present in /usr/gvx/bin/.......PASS
delay.esph.xyz.ZOOM file present in /usr/gvx/bin/.......PASS
Body B0 Image for scan 1 is less than 5 percent. PASS
Head B0 Image for scan 1 is less than 5 percent. PASS
Body B0 Image for scan 2 is less than 5 percent. PASS
Head B0 Image for scan 2 is less than 5 percent. PASS
Body B0 Image for scan 3 is less than 5 percent. PASS
Head B0 Image for scan 3 is less than 5 percent. PASS
Body B0 Image for scan 4 is less than 5 percent. PASS
Head B0 Image for scan 4 is less than 5 percent. PASS
Body B0 Image for scan 5 is less than 5 percent. PASS
Head B0 Image for scan 5 is less than 5 percent. PASS
Body B0 Image for scan 6 is less than 5 percent. PASS
Head B0 Image for scan 6 is less than 5 percent. PASS
Body B0 Image for scan 7 is less than 5 percent. PASS
Head B0 Image for scan 7 is less than 5 percent. PASS
Body B0 Image for scan 8 is less than 5 percent. PASS
Head B0 Image for scan 8 is less than 5 percent. PASS
**************************************************
--------------------------------------------------
PROBE Calibration File Checks
--------------------------------------------------
lrwxrwxr-x 1 33 Jan 20 17:28 /usr/g/caldir/probepfix.dat -> /usr/g/caldi
r/probepfix.dat.WHOLE%)
lrwxrwxr-x 1 33 Jan 20 17:28 /usr/g/caldir/probesfix.dat -> /usr/g/caldi
r/probesfix.dat.WHOLE
FAIL: probe tuning for probe-s and steamcsi all = 0.0. Probe tuning required.
steam_x =
steam_y =
steam_z =
press_x = -0.017241
press_y = 0.044486
press_z = -0.097118
**************************************************
--------------------------------------------------
Lcoil Calibration File Checks
--------------------------------------------------
WHOLE X Lcoil value in gram_tune.dat = 3500. Expected = 3500..................PA
SS
WHOLE Y Lcoil value in gram_tune.dat = 3200. Expected = 3200..................PA
SS
WHOLE Z Lcoil value in gram_tune.dat = 2900. Expected = 2900..................PA
SS
ZOOM X Lcoil value in gram_tune.dat = 3300. Expected = 3300..................PAS
S
ZOOM Y Lcoil value in gram_tune.dat = 3300. Expected = 3300..................PAS
S
ZOOM Z Lcoil value in gram_tune.dat = 3300. Expected = 3300..................PAS
S
**************************************************
--------------------------------------------------
Grafidy Calibration File Checks on WHOLE body
--------------------------------------------------
The tau ranges for X B0 eddy comp pass gross range check....PASS
The tau ranges for Y B0 eddy comp pass gross range check....PASS
The tau ranges for Z B0 eddy comp pass gross range check....PASS
The alpha ranges for X B0 eddy comp pass gross range check....PASS
The alpha ranges for Y B0 eddy comp pass gross range check....PASS
The alpha ranges for Z B0 eddy comp pass gross range check....PASS
The tau ranges for Linear on-axis eddy comp pass gross range check....PASS
The alpha ranges for Linear on-axis eddy comp pass gross range check....PASS
Cross term correction for X on Y is applied.
Cross term correction for X on Z is applied.
Cross term correction for Y on X is applied.
Cross term correction for Y on Z is applied.
Cross term correction for Z on X is applied.
Cross term correction for Z on Y is applied.
Cross term values for taus 3&4 and alphas 3&4 are zero....PASS
All applicable cross term tau values pass gross range check....PASS
X axis short time constant linear correction values enabled.
Y axis short time constant linear correction values enabled.
Z axis short time constant linear correction values enabled.
All applicable linear shorts alphas are positive....PASS
One or more of the VLTC taus or alphas have a non-zero value. VLTC's are enabled
.
--------------------------------------------------
Grafidy Calibration File Checks on ZOOM body
--------------------------------------------------
The tau ranges for X B0 eddy comp pass gross range check....PASS
The tau ranges for Y B0 eddy comp pass gross range check....PASS
The tau ranges for Z B0 eddy comp pass gross range check....PASS
The alpha ranges for X B0 eddy comp pass gross range check....PASS
The alpha ranges for Y B0 eddy comp pass gross range check....PASS
The alpha ranges for Z B0 eddy comp pass gross range check....PASS
The tau ranges for Linear on-axis eddy comp pass gross range check....PASS
The alpha ranges for Linear on-axis eddy comp pass gross range check....PASS
Cross term correction for X on Y is applied.
Cross term correction for X on Z is applied.
Cross term correction for Y on X is applied.
Cross term correction for Y on Z is applied.
Cross term correction for Z on X is NOT applied / not required.
Cross term correction for Z on Y is applied.
Cross term values for taus 3&4 and alphas 3&4 are zero....PASS
All applicable cross term tau values pass gross range check....PASS
X axis short time constant linear correction values enabled.
Y axis short time constant linear correction values enabled.
Z axis short time constant linear correction values enabled.
All applicable linear shorts alphas are positive....PASS
One or more of the VLTC taus or alphas have a non-zero value. VLTC's are enabled
.
**************************************************
--------------------------------------------------
T-tests File Checks
--------------------------------------------------
SPT nesting plate = Short
No TLT files found in /usr/gvx/service/data
No HSS files found in /usr/gvx/service/data
No RFT files found in /usr/gvx/service/data
The latest SST run = -rw-rw-r-- 1 21716 Nov 21 23:25 CBLN2524_WHOLE.SST
No RFS files found in /usr/gvx/service/data
The latest LVS run = -rw-rw-r-- 1 1288 Dec 13 05:41 CCD541RC_ZOOM.LVS
No NOT files found in /usr/gvx/service/data
The latest MCR run = -rw-rw-r-- 1 580 Dec 12 19:57 CCCJ5735.MCR
The latest FULL WHOLE SPT run = -rw-rw-r-- 1 364333 Dec 13 04:34 CCD43429_WH
OLE.SPT
The latest PARTIAL WHOLE SPT run = -rw-rw-r-- 1 362977 Jan 20 03:45 D1K34557
_WHOLE.SPTlth.log (93%)
The latest FULL ZOOM SPT run = -rw-rw-r-- 1 364333 Dec 13 05:25 CCD52547_ZOO
M.SPT
The latest PARTIAL ZOOM SPT run = -rw-rw-r-- 1 362973 Jan 20 04:10 D1K41044_
ZOOM.SPT
The above T test files can be viewed using REPORT TOOL or PC-REPORT.
Data from SPT File CCD43429_WHOLE.SPT
---------------------------------------
Number of days since SPT file CCD43429_WHOLE.SPT run = 40 days.
SPT file CCD43429_WHOLE.SPT Group 10 Noise = 45.25
SPT file CCD43429_WHOLE.SPT Group 10 Signal = 2256.89
SPT file CCD43429_WHOLE.SPT Group 10 SNR = 49.87
SPT file CCD43429_WHOLE.SPT Group 10 R1 = 11
SPT file CCD43429_WHOLE.SPT Group 10 R2 = 14
SPT file CCD43429_WHOLE.SPT Group 10 TG = 75
SPT file CCD43429_WHOLE.SPT Group 10 Cf = 116
SPT file CCD43429_WHOLE.SPT Group 10 Phan T1 = 63851051
SPT file CCD43429_WHOLE.SPT Group 79 Noise = 29.42
SPT file CCD43429_WHOLE.SPT Group 79 Signal = 966.59
SPT file CCD43429_WHOLE.SPT Group 79 SNR = 32.86
SPT file CCD43429_WHOLE.SPT Group 79 R1 = 11
SPT file CCD43429_WHOLE.SPT Group 79 R2 = 14
SPT file CCD43429_WHOLE.SPT Group 79 TG = 118
SPT file CCD43429_WHOLE.SPT Group 79 Cf = 134
SPT file CCD43429_WHOLE.SPT Group 79 Phan T1 = 63851206
The summary GROUP 9 from SPT file CCD43429_WHOLE.SPT is:
Z Isocenter PASS
Gradcal X PASS
Gradcal Y PASS
Gradcal Z PASS
Head Sys. Gain PASS
Head SNR PASS
Head FSE Stab. PASS
Head GRE Stab. PASS
Body Sys. Gain PASS
Body SNR PASS
Body FSE Stab. PASS
Body GRE Stab. PASS
Coherent Noise PASS
Data from SPT File D1K34557_WHOLE.SPT
---------------------------------------
Number of days since SPT file D1K34557_WHOLE.SPT run = 2 days.
Group 10 in SPT file D1K34557_WHOLE.SPT not found.
Group 79 in SPT file D1K34557_WHOLE.SPT not found.
The summary GROUP 9 from SPT file D1K34557_WHOLE.SPT is:
Head FSE Stab. PASS
Head GRE Stab. PASS
Body FSE Stab. PASS
Body GRE Stab. PASS
Data from SPT File CCD52547_ZOOM.SPT
---------------------------------------
Number of days since SPT file CCD52547_ZOOM.SPT run = 40 days.
SPT file CCD52547_ZOOM.SPT Group 10 Noise = 44.45
SPT file CCD52547_ZOOM.SPT Group 10 Signal = 2172.12
SPT file CCD52547_ZOOM.SPT Group 10 SNR = 48.87
SPT file CCD52547_ZOOM.SPT Group 10 R1 = 11
SPT file CCD52547_ZOOM.SPT Group 10 R2 = 14
SPT file CCD52547_ZOOM.SPT Group 10 TG = 76
SPT file CCD52547_ZOOM.SPT Group 10 Cf = 117
SPT file CCD52547_ZOOM.SPT Group 10 Phan T1 = 63851056
SPT file CCD52547_ZOOM.SPT Group 79 Noise = 35.02
SPT file CCD52547_ZOOM.SPT Group 79 Signal = 984.08
SPT file CCD52547_ZOOM.SPT Group 79 SNR = 28.10
SPT file CCD52547_ZOOM.SPT Group 79 R1 = 11
SPT file CCD52547_ZOOM.SPT Group 79 R2 = 14
SPT file CCD52547_ZOOM.SPT Group 79 TG = 122
SPT file CCD52547_ZOOM.SPT Group 79 Cf = 138
SPT file CCD52547_ZOOM.SPT Group 79 Phan T1 = 63851201
The summary GROUP 9 from SPT file CCD52547_ZOOM.SPT is:
Z Isocenter PASS
Gradcal X PASS
Gradcal Y PASS
Gradcal Z PASS
Head Sys. Gain PASS
Head SNR PASS
Head FSE Stab. PASS
Head GRE Stab. PASS
Body Sys. Gain PASS
Body SNR PASS
Body FSE Stab. PASS
Body GRE Stab. PASS%)
Coherent Noise PASS
Data from SPT File D1K41044_ZOOM.SPT
---------------------------------------
Number of days since SPT file D1K41044_ZOOM.SPT run = 2 days.
Group 10 in SPT file D1K41044_ZOOM.SPT not found.
Group 79 in SPT file D1K41044_ZOOM.SPT not found.
The summary GROUP 9 from SPT file D1K41044_ZOOM.SPT is:
Head FSE Stab. PASS
Head GRE Stab. PASS
Body FSE Stab. PASS
Body GRE Stab. PASS
**************************************************
--------------------------------------------------
Last LV shim Results
--------------------------------------------------
Magnet Type = GE LCC 1.5T
BW [Hz] = 200
FOV [cm] = 60.000000
DSV [cm] = 22.000000
Cf = 63851182
Inhomogeneity [Hz] = 4.225467
Inhomogeneity [ppm] = 0.066164
Z1=5.34 X=-3.61 Z2X=-1.09
Z2=-12.31 Y=-0.91 Z2Y=1.34
Z3=1.84 ZX=-1.99 Z(X2-Y2)=-2.85
Z4=3.44 ZY=-0.85 ZXY=2.67
Z5=-0.64 X2-Y2=4.02 X3=2.70
Z6=1.38 XY=-3.14 Y3=-1.45
**************************************************
--------------------------------------------------
Last CTL SNR DATA
--------------------------------------------------
USCS123 SNR DATA from file D1IK2859.CTL:
Coil=1 SNR=110.0 SNR std dev=0.232 PASS
Coil=2 SNR=64.3 SNR std dev=0.126 PASS
Coil=3 SNR=95.5 SNR std dev=0.366 PASS
Coil=1 L-R Top Shading = 0.989
Coil=1 L-R Bot Shading = 0.796
Coil=1 Top-Bot Shading = 0.655
Coil=1 Xtalk1 = 0.336
Coil=1 Xtalk2 = 0.311
Coil=1 Xtalk3 =
Coil=2 L-R Top Shading = 1.257
Coil=2 L-R Bot Shading = 1.107
Coil=2 Top-Bot Shading = 0.303
Coil=2 Xtalk1 = 0.099
Coil=2 Xtalk2 = 0.135
Coil=2 Xtalk3 =
Coil=3 L-R Top Shading = 1.092
Coil=3 L-R Bot Shading = 1.007
Coil=3 Top-Bot Shading = 0.276
Coil=3 Xtalk1 = 0.213
Coil=3 Xtalk2 = 0.162
Coil=3 Xtalk3 =
CS123 SNR DATA from file D1IG4910.CTL:
Coil=1 SNR=166.2 SNR std dev=81.422 FAIL
Coil=2 SNR=646.1 SNR std dev=316.546 PASS
Coil=3 SNR=807.3 SNR std dev=395.520 PASS
**************************************************
--------------------------------------------------
Last MCR Results from file CCCJ5735.MCR
--------------------------------------------------
Rec#=1 Sig=1214.0 Noise=14.33 SNR=84.7 Spec=81.3 ...........PASS
Rec#=2 Sig=1378.1 Noise=16.16 SNR=85.3 Spec=81.3 ...........PASS
Rec#=3 Sig=1207.6 Noise=14.31 SNR=84.4 Spec=81.3 ...........PASS
Rec#=4 Sig=1211.6 Noise=14.39 SNR=84.2 Spec=81.3 ...........PASS
Rec#=5 Sig=1291.0 Noise=15.18 SNR=85.0 Spec=81.3 ...........PASS
Rec#=6 Sig=1379.3 Noise=16.17 SNR=85.3 Spec=81.3 ...........PASS
Rec#=7 Sig=1363.1 Noise=15.97 SNR=85.4 Spec=81.3 ...........PASS
Rec#=8 Sig=1459.6 Noise=16.94 SNR=86.2 Spec=81.3 ...........PASS
**************************************************
--------------------------------------------------
EPI White Pixel results files check
--------------------------------------------------
WHOLE BODY EPIWP RESULTS
Date/time= 12/13/02,05:50:37
Operator= 00000
Bay= b5
Reli= 0
Process= A
Iteration= 1
Exam= 51342
BRM unit number= 0
Noise Mean Upper Limit= 15.8
Noise Mean Lower Limit= 4.8
Noise SD Upper Limit= 8.4
Noise SD Lower Limit= 2.6
system_health.log (96%)
ZOOM BODY EPIWP RESULTS
Date/time= 12/13/02,07:49:00
Operator= 00000
Bay= b5
Reli= 0
Process= A
Iteration= 1
Exam= 51356
BRM unit number= 0
Noise Mean Upper Limit= 15.8
Noise Mean Lower Limit= 4.8
Noise SD Upper Limit= 8.4
Noise SD Lower Limit= 2.6
Img 1 fails 4 WP, axis=x, mean=9.860, SD=5.621, th=194.1, clus=3, max=296
Img 3 fails 45 WP, axis=z, mean=9.860, SD=5.621, th=194.1, clus=27, max=584
Img 4 fails 2 WP, axis=x, mean=9.551, SD=5.164, th=181.2, clus=1, max=325
Img 6 fails 2 WP, axis=z, mean=9.551, SD=5.164, th=181.2, clus=2, max=204
Img 9 fails 28 WP, axis=z, mean=9.989, SD=4.944, th=178.6, clus=22, max=678
Img 12 fails 2 WP, axis=z, mean=9.633, SD=5.343, th=186.0, clus=2, max=191
Img 15 fails 3 WP, axis=z, mean=10.563, SD=5.404, th=193.1, clus=3, max=313
Img 16 fails 2 WP, axis=x, mean=10.477, SD=5.565, th=196.4, clus=2, max=303
Img 18 fails 14 WP, axis=z, mean=10.477, SD=5.565, th=196.4, clus=11, max=415
Img 19 fails 3 WP, axis=x, mean=11.046, SD=5.677, th=202.5, clus=3, max=238
Img 21 fails 5 WP, axis=z, mean=11.046, SD=5.677, th=202.5, clus=4, max=361
Img 24 fails 24 WP, axis=z, mean=10.132, SD=5.311, th=188.3, clus=16, max=802
Img 30 fails 2 WP, axis=z, mean=10.449, SD=5.706, th=199.6, clus=2, max=235
Img 31 fails 1 WP, axis=x, mean=10.335, SD=5.176, th=186.2, clus=1, max=209
Img 33 fails 10 WP, axis=z, mean=10.335, SD=5.176, th=186.2, clus=8, max=396
Img 36 fails 12 WP, axis=z, mean=10.621, SD=5.065, th=185.3, clus=10, max=346
Img 39 fails 9 WP, axis=z, mean=10.247, SD=5.465, th=192.6, clus=8, max=341
Img 42 fails 2 WP, axis=z, mean=9.991, SD=5.682, th=196.3, clus=1, max=294
Img 45 fails 6 WP, axis=z, mean=10.367, SD=5.841, th=202.4, clus=5, max=427
Img 48 fails 1 WP, axis=z, mean=10.752, SD=5.661, th=200.4, clus=1, max=378
File sn_report.head.WHOLE not in /usr/g/bin. WHOLE Head epiwp data not available
.
File sn_report.head.ZOOM not in /usr/g/bin. ZOOM Head epiwp data not available.
**************************************************
--------------------------------------------------
Free disk block report (Unix df command).
--------------------------------------------------
Filesystem Type blocks use avail %use Mounted on
/dev/root xfs 174720 62232 112488 36 /
/dev/dsk/dks0d1s6 xfs 5581440 4790512 790928 86 /usr
/dev/dsk/dks0d1s5 xfs 2459024 1734656 724368 71 /export/home
/dev/dsk/dks0d1s7 xfs 8192640 6823872 1368768 84 /export/home1
**************************************************
--------------------------------------------------
Option key status.
--------------------------------------------------
cine INSTALLED
epli INSTALLED
fastgrass INSTALLED
fse2 INSTALLED
pcvi INSTALLED
research INSTALLED
spectro INSTALLED
proactive INSTALLED
gtof INSTALLED
resrchpsd INSTALLED
vconf INSTALLED
sgd INSTALLED
dwepi INSTALLED
flairepi INSTALLED
special INSTALLED
smartprep INSTALLED
ssfse INSTALLED
threeplane INSTALLED
hisris INSTALLED
e3dtof INSTALLED
fsexl INSTALLED
bloodsupp INSTALLED
fastcine INSTALLED
tagging INSTALLED
sgdperf INSTALLED
dwepiez INSTALLED
idrivePro INSTALLED
idrive INSTALLED
mns INSTALLED
sprep99 INSTALLED
probe99 INSTALLED
sage INSTALLED
pps INSTALLED
hispeed INSTALLED
perfusion INSTALLED
rtet INSTALLED
spiralHR INSTALLED
spiralRT INSTALLED
rtca INSTALLED
probe3d INSTALLED
ushorttr INSTALLED
t2bhold INSTALLED
ssfsemrcp INSTALLED
t1bhold INSTALLED
LSDI INSTALLED
ifilter INSTALLED
acgdplus INSTALLED
prose INSTALLED
ftmra INSTALLED
ssfsexl INSTALLED
psir INSTALLED
DynR1 INSTALLED
fiesta2d INSTALLED
fiesta3d INSTALLED
delenhmt INSTALLED
asset INSTALLED
probep INSTALLED
tensor INSTALLED
FR3DFSE INSTALLED
OPTION@@@STATUS
------@@@------
ivi@@@INSTALLED
vxtl@@@INSTALLED
fctl2@@@INSTALLED
clariview@@@INSTALLED
**************************************************
--------------------------------------------------
Host hardware inventory
--------------------------------------------------
1 360 MHZ IP30 Processor
CPU: MIPS R12000 Processor Chip Revision: 3.5
FPU: MIPS R12010 Floating Point Chip Revision: 0.0
Main memory size: 512 Mbytes
Xbow ASIC: Revision 1.4
Instruction cache size: 32 Kbytes
Data cache size: 32 Kbytes
Secondary unified instruction/data cache size: 2 Mbytes
Integral SCSI controller 0: Version QL1040B (rev. 2), single ended
Disk drive: unit 1 on SCSI controller 0
Integral SCSI controller 1: Version QL1040B (rev. 2), single ended
CDROM: unit 3 on SCSI controller 1
Optical disk: unit 4 on SCSI controller 1
Integral SCSI controller 2: Version QL1040B (rev. 2), single ended
Comm device: unit 4 on SCSI controller 2
Comm device: unit 4, lun 1 on SCSI controller 2
IOC3 serial port: tty1
IOC3 serial port: tty2
IOC3 parallel port: plp1
Graphics board: V10
Integral Fast Ethernet: ef0, version 1, pci 2
Fast Ethernet: ef1, version 1, pci 1
Iris Audio Processor: version RAD revision 12.0, number 1
PCI card, bus 0, slot 0, Vendor 0x108a, Device 0x0
PCI card, bus 0, slot 224, Vendor 0x108a, Device 0x0
PCI card, bus 0, slot 32, Vendor 0x108a, Device 0x0
PCI card, bus 0, slot 64, Vendor 0x108a, Device 0x0
PCI card, bus 0, slot 160, Vendor 0x108a, Device 0x0
PCI card, bus 0, slot 96, Vendor 0x108a, Device 0x0
PCI card, bus 0, slot 128, Vendor 0x108a, Device 0x0
PCI card, bus 0, slot 192, Vendor 0x108a, Device 0x0
**************************************************
--------------------------------------------------
The following files from product group MrpApps have
checksums different than the original LFC checksums.
This indicates these files have been patched or
changed from the original LFC versions of the files.
--------------------------------------------------
f 64537 5176 MrpApps.sw.base usr/g/bin/ept/gep/other/1.EPT
f 19174 5180 MrpApps.sw.base usr/g/bin/ept/gep/other/2.EPT
f 20983 5181 MrpApps.sw.base usr/g/bin/ept/gep/other/3.EPT
f 25783 5189 MrpApps.sw.base usr/g/bin/ept/gep/other/4.EPT
f 31355 5188 MrpApps.sw.base usr/g/bin/ept/gep/other/5.EPT
f 30635 5187 MrpApps.sw.base usr/g/bin/ept/gep/other/6.EPT
f 3141 5193 MrpApps.sw.base usr/g/bin/ept/gep/other/7.EPT
f 10404 5191 MrpApps.sw.base usr/g/bin/ept/gep/other/8.EPT
f 7494 132 MrpApps.sw.base usr/g/hisris/WLFilterFile
f 14151 18008 MrpApps.sw.base usr/g/insite/bin/grablock
f 35368 3048 MrpApps.sw.base usr/g/insite/custom/top.conf.local
f 33950 4378 MrpApps.sw.base usr/g/protocols/service.1.5T/other/Grafid
y
f 45165 14025179 MrpApps.sw.base usr/g/spectro/rsi/idlsgi511rt.tar.Z
f 64923 212739 MrpApps.sw.base usr/gvx/bin/spu.out
f 24718 76 MrpApps.sw.base usr/gvx/caldir/grafidyx.sgd.ssn
f 24718 76 MrpApps.sw.base usr/gvx/caldir/grafidyy.sgd.ssn
f 30862 76 MrpApps.sw.base usr/gvx/caldir/grafidyz.sgd.ssn
**************************************************
--------------------------------------------------
MGD UTNS count mode and counts
--------------------------------------------------
**************************************************
*********** END OF SYSTEM HEALTH DATA **********