Removing Active Shim Coils
Overview
The ShimConfig.cfg file is located in /usr/g/w/config and contains the following information for each coil:
-
harmonic naming order
-
current calculation; on or off
-
existing coil current value
-
pertubation current value
-
maximum power supply current
-
power supply or gradient coil name
-
harmonic name.
The vi editor is used to edit this file.
Coil Removal
-
At the operator workspace, select the Tools icon in the desktop control panel.
-
In the Service Browser, select the Calibration tab, then in the menu options on the left side of the browser, click LVShim. On the right side, click the link Click here to start this tool. Refer to Table 1.
| {unix prompt} .................................. | cd /usr/g/w/configEnter |
| {unix prompt} .................................. | Verify the file exists ls -al Shim*.*Enter ShimConfig.cfg is part of Load From Cold software. |
|
|
|