• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30877E20GYZ
  • Topic ID: id_13106204 Version: 3.0
  • Date: Aug 29, 2019 1:36:46 AM

CRM TRMZM Shim Specs

Monitor the overall inhomogeneity change from one iteration to the next. If you are seeing no more than a 1-Hz change between iterations, stop shimming. If you are unable to shim the magnet below the specified overall inhomogeneity, you may need to:

  • Check and adjust B0. Refer to the Eddy Current Compensation procedure for details on measuring and adjusting B0.

  • Check for image artifacts, such as zippers, spike noise, etc.

  • Call your Zone Support Engineer or OLC for assistance if the system cannot achieve the magnet shim specifications.

Note:

Most magnets should be able to achieve the “Shim To” values shown. All magnets must meet the “Customer Acceptance” values. As long as the magnet meets the Customer Acceptance values, Gradient LVshim should be sufficient to maintain an acceptable shim quality. When the Customer Acceptance values are exceeded, Main LVshim must be performed to re-establish an acceptable shim quality.

Note:

All Harmonics are Absolute Values and given as Shim to Goal / Customer Acceptance.

See Table 1 for 3.0T Specs, Table 2 for 1.5T Specs or Table 3 for 1.0T Specs.

Table 1. 3.0T
DSV = 40 cm
Values are shown as: Shim to Goal / Customer Acceptance
Magnet Type
Harmonic Name 3T94 Eclipse 2 With EXCITE 3T94 Eclipse
BW400 Hz1000 Hz
STD DEV Hz<45/45
STD DEV PPM0.35
1,0 (Z1)<14/25<10/100
2,0 (Z2)<25/60<25/100
3,0 (Z3)<50/50<25/100
4,0 (Z4)<100/100<25/100
5,0 (Z5)<100/100<127/240
6,0 (Z6)<200/200<127/240
1, -1 (X)<14/25<10/100
1, 1 (Y)<14/25<10/100
2, -1 (ZX)<50/50<50/50
2, 1 (ZY)<50/50<25/50
2, -2 (X2-Y2)<50/50<25/50
2, 2 (XY)<50/50<25/50
3, -1 (Z2X)<25/50<25/50
3, 1 (Z2Y)<25/50<25/50
3, -2 (Z(X2-Y2))<25/50<25/50
3, 2 (ZXY)<25/50<25/50
3, -3 (X3)<100/100<127/180
3, 3 (Y3)<100/100<127/180
Table 2. 1.5T
DSV = 45 cm
BW = 500 Hz
Values are shown as: Shim to Goal / Customer Acceptance
Magnet Type
S1 (14 Coils) SII, SIII, SIV SV (12 Coil) CX, LCC
STD DEV Hz <50/64<50/64<50/64<50/64
STD DEV PPM <0.6/1.00<0.6/1.00<0.6/1.00<0.6/1.00
Harmonic Name
1,0 (Z1) 1 count1 count1 count1 count
2,0 (Z2) <25/50<25/50<25/50<25/50
3,0 (Z3) <50/100<50/100<50/100<25/50
4,0 (Z4) <50/100<50/100<50/100<50/100
5,0 (Z5) <10/100<50/100<50/100<50/100
6,0 (Z6) <200/200<100/200<100/200<50/100
1, -1 (X) 1 count1 count1 count1 count
1, 1 (Y) 1 count1 count1 count1 count
2, -1 (ZX) <25/50<25/50<25/50<25/50
2, 1 (ZY) <25/50<25/50<25/50<25/50
2, -2 (X2-Y2) <50/100<50/100<100/100<25/50
2, 2 (XY) <50/100<50/100<100/100<25/50
3, -1 (Z2X) <25/50<25/50<25/50<25/50
3, 1 (Z2Y) <25/50<25/50<25/50<25/50
3, -2 (Z(X2-Y2)) <25/50<25/50<25/50<25/50
3, 2 (ZXY) <25/50<25/50<50/50<25/50
3, -3 (X3) <100/100<50/100<100/100<25/50
3, 3 (Y3) <100/100<50/100<100/100<25/50
Table 3. 1.0T
DSV = 45 cm
BW = 500 Hz
Values are shown as:

Shim to Goal / Customer Acceptance

Magnet Type
SX SX, SXC, SXCII CX, LCC
STD DEV Hz<33/43<33/43<33/43
STD DEV PPM<.06/1.00<.06/1.00<.06/1.00
Harmonic Name
1,0 (Z1)1 count1 count1 count
2,0 (Z2)<25/50<25/50<25/50
3,0 (Z3)<50/100<50/100<25/50
4,0 (Z4)<50/100<50/100<50/100
5,0 (Z5)<50/100<50/100<50/100
6,0 (Z6)<100/200<200/200<50/100
1, -1 (X)1 count1 count1 count
1, 1 (Y)1 count1 count1 count
2, -1 (ZX)<25/50<25/50<25/50
2, 1 (ZY)<25/50<25/50<25/50
2, -2 (X2-Y2)<50/100<100/100<25/50
2, 2 (XY)<50/100<100/100<25/50
3, -1 (Z2X)<25/100<25/50<25/50
3, 1 (Z2Y)<25/50<25/50<25/50
3, -2 (Z(X2-Y2))<25/50<50/50<25/50
3, 2 (ZXY)<25/50<50/50<25/50
3, -3 (X3)<50/100<10/100<25/50
3, 3 (Y3)<50/100<100/100<25/50