Monitor the overall inhomogeneity change from one iteration to the next. If you are seeing no more than a 1-Hz change between iterations, stop shimming. If you are unable to shim the magnet below the specified overall inhomogeneity, you may need to:

  • Check and adjust B0. Refer to the Eddy Current Compensation procedure for details on measuring and adjusting B0.

  • Check for image artifacts, such as zippers, spike noise, etc.

  • Call your Zone Support Engineer or OLC for assistance if the system cannot achieve the magnet shim specifications.

Note:

Most magnets should be able to achieve the “Shim To” values shown. All magnets must meet the “Customer Acceptance” values. As long as the magnet meets the Customer Acceptance values, Gradient LVshim should be sufficient to maintain an acceptable shim quality. When the Customer Acceptance values are exceeded, Main LVshim must be performed to re-establish an acceptable shim quality.

Note:

All Harmonics are Absolute Values and given as Shim to Goal / Customer Acceptance

See Table 1 for 3.0T Specs, Table 2 for 1.5T.

Table 1. 3T
DSV = 45 cm
BW = 400 Hz
Values are shown as: Shim to Goal / Customer Acceptance
Magnet Type
LCC300
STD DEV Hz <64/127
STD DEV PPM 0.5/1.00
Harmonic Name
1,0 (Z1) <12/25
2,0 (Z2) <25/60
3,0 (Z3) <25/50
4,0 (Z4) <50/100
5,0 (Z5) <100/100
6,0 (Z6) <200/200
1, -1 (X) <12/25
1, 1 (Y) <12/25
2, -1 (ZX) <25/50
2, 1 (ZY) <25/50
2, -2 (X2-Y2) <25/50
2, 2 (XY) <25/50
3, -1 (Z2X) <25/50
3, 1 (Z2Y) <25/50
3, -2 (Z(X2-Y2)) <25/50
3, 2 (ZXY) <25/50
3, -3 (X3) <100/100
3, 3 (Y3) <100/100
Table 2. 1.5T
DSV = 45 cm
BW = 200 Hz
Values are shown as: Shim to Goal / Customer Acceptance
Magnet Type
Oxford (12 Coils)S1 (14 Coils)SII, SIII, SIVSV (12 Coil)CX, LCC
STD DEV Hz<80/80<50/80<50/80<50/80<50/80
STD DEV PPM<1.25/1.25<0.6/1.25<0.6/1.25<0.6/1.25<0.6/1.25
Harmonic Name
1,0 (Z1)<10/35<10/35<10/35<10/35<10/35
2,0 (Z2)<25/50<25/50<25/50<25/50<25/50
3,0 (Z3)<50/100<50/100<50/100<50/100<25/50
4,0 (Z4)<50/100<50/100<50/100<50/100<50/100
5,0 (Z5)<50/100<100/100<50/100<50/100<50/100
6,0 (Z6)<100/200<200/200<100/200<100/200<50/100
1, -1 (X)<10/35<10/35<10/35<10/35<10/35
1, 1 (Y)<10/35<10/35<10/35<10/35<10/35
2, -1 (ZX)<50/100<25/50<25/50<25/50<25/50
2, 1 (ZY)<50/100<25/50<25/50<25/50<25/50
2, -2 (X2-Y2)<50/100<50/100<50/100<100/100<25/50
2, 2 (XY)<50/100<50/100<50/100<100/100<25/50
3, -1 (Z2X)<50/100<25/50<25/50<25/50<25/50
3, 1 (Z2Y)<50/100<25/50<25/50<25/50<25/50
3, -2 (Z(X2-Y2))<100/100<25/50<25/50<50/50<25/50
3, 2 (ZXY)<100/100<25/50<25/50<50/50<25/50
3, -3 (X3)<100/100<100/100<50/100<100/100<25/50
3, 3 (Y3)<100/100<100/100<50/100<100/100<25/50