Running LVShim - Gradshim
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 30 minutes | 10 minutes |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| LVShim Phantom Inner Sphere | 1 | 2274476 | - |
| LV Phantom Small Half Assembly | 1 | 2306642 | - |
| LV Phantom Big Half Assembly | 1 | 2306645 | - |
| Curved Adapter Panel for Flat Table (GEM) | 1 | 5395828 | - |
| Universal Phantom Pad | 1 | 46-328383P1 | - |
| Service Filler Panel | 1 | 5344577 | - |
| Required conditions |
|---|
| Confirm that passive shimming is complete and is within specification. |
| Remove all coils from the cradle and phantoms from the bore. |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
The Gradshim process is done with a sampling diameter of 22 cm DSV and the same bandwidth as LVShim Fine, which depends on the quality (good/bad) of the fine tuning. Once the scan passes and all harmonics are within specification, the magnetic field is reanalyzed in Test mode automatically at 45 DSV for customer acceptance.
The gradshim process includes LVShim - test. The objective of this test is to analyze completed scans at 45 DSV without calculating currents.
- On the MR Service Desktop, select the PM tab.
- Select .
| Running LVShim - Gradshim for systems running DV26.0 or earlier |
| Running LVShim - Gradshim for systems running DV29.1 or later |