Finalizing Grafidy calibration
Procedure
- Exit the software tool, saving the eddy current data.
- Remove the Grafidy fixture from the system.
- If calibration is being run within ICW installation mode continue to the next calibration procedure. Otherwise, Save Grafidy eddy current data to SaveInfo media. See Saving information.