Calibrating grafidy
Grafidy adjusts and compensates short, long, and very long time constant eddy currents for both linear and B0 terms. Use the Grafidy procedure to do eddy current compensation.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 2 hours | 5 minutes |
| Tools and test equipment | ||||
|---|---|---|---|---|
| Item | Quantity | Part number | Manufacturer | |
| 1.5T Grafidy Kit | 1 | - | 5135527-7 | |
| Service Filler Panel | 1 | - | 5344577 | - |
| Service Foam Positioner | 1 | - | 5554839 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| RF coil assembly | 1 | 5135514 | - |
| Required conditions |
|---|
| DQA (Grad Cal/IsoVector2) tool within specification. See DQA calibration. |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
Grafidy uses a dedicated pulse sequence, six-channel fixture, and Grafidy software. The Grafidy P-port connector plugs into the port for transmit/receive. No additional fixtures or cables are needed.
This tool simplifies setting up the calibration and collects data from all six coils simultaneously.
| Auto mode - The Grafidy tool is initially run in auto mode. |
| Manual mode - When doing this procedure, if auto mode does not bring all the terms into specification (still red in the Result/Area table). Note: If VLONG, LONG or SHORT terms are failing, manual mode should be run to bring the VLONG mode into specification. |