• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30C11E20GYZ
  • Topic ID: id_13106417 Version: 7.0
  • Date: Feb 21, 2021 9:08:02 PM

UPM Calibration and UPM Functional Check–Head Mode

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1 -30 minutes (plus 30 minutes for Agilent wattmeter power up)10 minutes
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Medium Phillips screwdriver1---
12 inch crescent wrench1---
RF power measurement kit (one of the following options):1-

EITHER 5307511-2 or 5307511-3 (Bird wattmeter)

BOTH 5434817 (Agilent wattmeter) and 5434817-20 (dummy load)

-
Table 3. Safety
Table 4. Required conditions
ConditionReferenceEffectivity

When using the Agilent wattmeter, plug in the wattmeter and turn it on. The wattmeter must be powered for at least 30 minutes for stable and accurate readings.

--

About this task

Use the following procedures to calibrate the head universal power monitor (UPM) and perform the functional check for the head UPM. These procedures must be performed in sequence:

  • The UPM calibration adjusts digital attenuator values internal to the UPM to match the expected RF signal levels.

    Calibration is performed for forward power and reflected power.

  • The UPM functional check runs a prescribed scan in the forward power. It compares the measured RF power from the UPM to the expected value stored in the UPM configuration file. If two values are outside 12%, the functional check fails. If the measured RF power exceeds the trip limit, the UPM stops scanning.

Dual drive RF amplifier systems operate in average power mode. Functional check tests include: 10-second average power check, RF inhibit test, and short-term average power fault host trip test.

Table 5. Calibration Steps: Dual Drive Amplifier
Calibration Steps
Safety
Head Disabling T/R, DD, and RF Input for Head Channel
Hardware Setup for RF Head Calibration
UPM Calibration for Head Channel
UPM Functional Check for Head Channel
Finalization

If you are performing an initial installation and need to calibrate the RF amplifier, calibrate the UPMs, and check the UPMs, use the procedures in RF Amp Body and Head Maximum Power Setup and Calibration.

Topic ID: id_SL1753389-1213179

Disabling T/R, DD, and RF Input for Head Channel

Procedure

  1. (For Agilent wattmeters) Plug in the wattmeter and turn it on. The wattmeter must be powered for at least 30 minutes for stable and accurate readings.
  2. Verify that the system is idle.
  3. Open the front door of the ISC.
  4. On the front of the driver module, disable the TR and DD fault detection.
    1. Set SW2 to TR DISABLE (down) and set SW3 to DD DISABLE (down).
      Note: Leave HV set to ENABLE.
    2. Set RF Enable to DISABLE (down).
    Figure 1. Disabling RF, TR, and DD
Topic ID: id_SL12187137-1213179

UPM Calibration for Head Channel

Procedure

  1. Set RF ENB to ENABLE (up).
    Figure 2. RF ENB
  2. Press Landmark. Then, advance to scan. A phantom is not needed.
    Note: No phantom is needed since dummy command for head connection is set in UPM Tool Head mode for SIGNA Pioneer.
    Figure 3. Adv to scan
  3. (For non-proprietary mode) On the Common Service Desktop, select Calibration > UPM Tool and select Click here to start this tool.
  4. Set Calibrating the UPM in Head mode.
    Figure 4. UPM Calibration Settings for Head Forward
    For this setting: Select:
    Nuclei 1H
    Task Calibrate UPM
    RF Chain Head
    Power Type Forward
  5. Select Start.
  6. Select Yes when the prompt asks if the hardware setup is correct.

    After approximately one minute, you are asked to confirm the peak power value output from the amplifier.

    If the power read from the power meter is within specification, click Yes. Otherwise, click No and recheck the calibration of the RF amplifier.

  7. The tool reports the steps as they progress. In about three minutes, the Status field on the UPM Tool displays PASS or FAIL.

    A popup reminds you that you must run reflected power after you have run forward power.

  8. After calibrating forward power, calibrate reflected power as follows:
    1. Set RF ENB to DISABLE (down).
    2. On the front of the RF amplifier, disconnect the coupler from HEAD RF OUT/J3. Leave J3 open and unconnected.
    3. Set RF ENB to ENABLE (up).
    4. From the UPM Tool, make the following selections:
      Figure 5. UPM Calibration Settings for Head Reflected
      For this setting: Select:
      Nuclei 1H
      Task Calibrate UPM
      RF Chain Head
      Power Type Reflected
    5. Select Start.
    6. Select Yes when the prompt asks if the hardware is set up correctly.
    7. Upon completion of reflected power calibration, return the cable configuration to support testing for forward power.
      1. Set RF ENB to DISABLE (down).

      2. On the front of the RF amplifier, connect the coupler to HEAD RF OUT/J3.

      3. Set RF ENB to ENABLE (up).

Topic ID: id_SL12187141-1213179

UPM Functional Check for Head Channel

Procedure

  1. After completing the UPM calibration, continue with the UPM functional check.
    1. On the UPM Tool, make the following selections:
      Figure 6. UPM Functional Check Settings for Head
      For this setting: Select:
      Nuclei 1H
      Task UPM Func Chk
      RF Chain Head
      Note: Power Type is grayed out. This is normal and is not used.
    2. Select Start.
    3. Select YES when the prompts ask if the hardware is set up correctly.
    4. Prompts are displayed during test execution, including a pop up message about the UPM tripping and a power trip error log.
      Note: The tool performs three tests (10-second average power, UPM1 RF inhibit, and UPM2 RF inhibit) before host trip. This takes 2-3 minutes before user interaction is required.
      Figure 7. UPM Power Trip Error Log
    5. When PASS appears on the UPM Tool interface, check that all the body channel is calibrated and checked.
  2. Restore RF cables per following steps.
    1. Set the RF ENB switch to DISABLE (down).
    2. Disconnect the RF coupler from J3 and reconnect the head and transmit cables.
    3. Re-enable the TR and DD fault switches on the front of the driver module.
    4. Set RF ENB switch to ENABLE (up).
    Figure 8. Restore RF Amp
Topic ID: id_15013039

Finalization

Procedure

  1. Complete a TPS reset and perform Check Scan.
  2. If all the calibrations have been successfully made, perform a SaveInfo to save the UPM calibration results.