Doing the power supply data diagnostic

This diagnostic reads the power supply information from the Driver Control Board (DCB) board every five seconds for the specified test duration (maximum of 300 seconds).

Topic ID: id_SL11658706-1224765

Disclaimer

Depending on your service agreement, not all service tools, diagnostics, and utilities referenced in this document may be accessible. Contact your sales person for information on available service license packages.

Topic ID: id_SL11658707-1224765

Diagnostic description

Topic ID: id_SL1943696-1224765

Diagnostic path

Diagnostics > System Function > RF > Driver Module Diagnostics

Diagnostics > Hardware Location > Integrated System Cabinet > RF > Driver Module Diagnostics

Topic ID: id_SL1943698-1224765

Purpose

This diagnostic reads the power supply information from the DCB board every five seconds for the specified test duration (maximum of 300 seconds). The values are checked against the tolerances listed in the DcbConfig.cfg file.

Topic ID: id_SL1943699-1224765

Components tested

  • Driver module
Topic ID: id_SL1943700-1224765

Test sequence

  1. Select Power Supply Data and adjust the settings as needed, including the test duration.
  2. Click Run.
Topic ID: id_SL1943701-1224765

Results

Test results

Diagnostic results display a status of Pass or Fail. If you receive a Fail status, click GE System Log to view the error message. The system log records errors with the date, time, the process tested, and a detailed error message.

Test status

This window informs you of the progress of the test from the initial scan to the completed diagnostic.