- Discovery MR750 3.0T System Service Methods
- 5690009-2EN Revision 4
- 00000018WIA30BCBE20GYZ
- id_131064943.4
- Oct 11, 2021 6:13:21 PM
Body coil air flow functional check
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 10 to 15 minutes | Not Applicable |
| ||||
| Condition | Reference | Effectivity |
|---|---|---|
|
System software must be booted. | - | - |
About this task
This test checks the functionality of the body coil air blower system, which cools the patient bore wall surfaces by forcing air between the gradient coil and the RF coil Figure 1. Air enters the magnet through two rear air ports; therefore, the rear of the magnet needs to have an airtight seal in order to maintain a sufficient pressure head to force air to flow. This helps to transfer heat away from the bore wall with convection. If there is not an air tight seal in the rear of the magnet, there will not be enough pressure to force air through the space between the gradient and RF coils.

The block diagram below shows the system configuration. Air is pulled out of the magnet room, and travels through the penetration wall and into the heat exchange cabinet (HEC). This air is then cooled and sent back through the penetration wall and into the magnet room. This cooled air then travels from the penetration wall, splits into two air hoses, and into the back of the magnet through tubing in the rear end bell or through tubing in the body coil. At the front of the magnet are two air sensors that capture an average airflow measurement and relay the information back to the HEC display. This airflow requires a measurement of greater than 500 feet per minute (fpm) during scanning operations. If the airflow measurement is less than 500 fpm, see Troubleshooting body coil air flow.

Checking body coil air flow
About this task
This procedure consists of taking readings from the HEC display on the heat exchanger cabinet and determining if an error is occurring with the air flow system.
Procedure
- Go to the heat exchanger cabinet and view the display.
- Select the following buttons from the main menu display in order
to view the airflow set point and feedback readings:
- Note: Select ESC multiple times to return to the main menu display.Monitor (F3)
- Blower (F4)
- Confirm that the airflow set point corresponds to a measurement equal to or greater than 550 fpm.
- Confirm that the air flow feedback corresponds to a measurement
greater than 500 fpm.
- If the feedback is lower than 500 fpm, see Troubleshooting body coil air flow to troubleshoot the body coil air flow.
- If the feedback is greater than or equal to the Normal mode set point, the system is functioning properly.
Troubleshooting body coil air flow
About this task
Procedure
Finalization
Finalization
No finalization steps.











