System Cable FRU

Table 1. System Cable FRU
Part Number Description
5477605-2Run P2003 ISC-ICE-J1, 2, 5 to MAG-DPP1 J6, 9, DPP2 J6, 17050 mm
5477605-4Run P2004 ISC-ICE-J6, 13, 14 to MAG-DPP2 J9, DPP3 J6, 9, 17700 mm
5477605-8Run P2006 ISC-ICE-J30 to MAG-RXDIST J16, Spares, 17800 mm
5476645-2Run P2009 MAG-PAC J1 to ISC-ICE-J38, J39, CAN Fiberoptic, 16550 mm
5476646-2Run P2010 MAG-SRI J11_J12 to ISC-ICE-J40_J41, CAN Fiberoptic, 17400 mm
5476646-3Run P2011 MAG-SRI J13 J14 to ISC-PHPS2-J37_J38, CAN Fiberoptic, 17150 mm
5475817-2Run M3311 ISC-PHPS2-J36 to Table IO-J5, D-sub, Long Drive Power, Nonmagnetic, 18500 mm
5476366-2Run M3315 ISC-PHPS2-J54 to Table IO-Table Cont Box-J4, Halls Signal, Nonmagnetic, 18500 mm
5476816 Run M3314, Cable IRD_OCP power, Nonmagnetic
5475829-2Run M3313 ISC-PHPS2-J35 to Table IO-Table Cont Box-J8, TABLE POWER, Non Magnetic, 18500 mm
5476352-3Run M1302, ISC-PW-J29 to RxDist-J19, Receiver Clock, Twinax, 15,140 mm
5476352-4Run M1304, ISC-PW-J30 to MAG-RxDist-J10, Rx LO, Twinax, 16,330 mm
5476371-2Run M3316 ISC-PW-J17 to MAG-Rx Dist-J18, Inhibit, Nonmagnetic, 15140 mm
5476630-2Run M1303, MAG-RxDist-J12 to ISC-PW-J32, Loopback, 15,550 mm
5476767-2Run M3384 ISC-PW-J75 TO MAG-DTRSW J7, 15000 mm
5476767-44Run M3385 ISC-PW-J76 TO MAG-DTRSW J8, 15000 mm
5476636-2Run M3300 ISC-PW-J112 to MAG-Rx Dist J1, Rx Dist Power, 24W7, Low-Mag, 15750mm
5476751-2Run M3302 MAG-SRI-J1 to ISC-PHPS2-J40,SRI Power,Nonmagnetic, 15000 mm
5475840-2Run M3303 ISC-PHPS2-J39 to MAG-SRI J21, MCPDB Interface, Nonmagnetic, 15000 mm
5475830 Run M3360 ISC-PHPS2-J41 to MAG-PAC-J2, MAG-PAC-J9 Y-Cable DV PAC for Power and One-Wire,15000 mm
5476796 Run M1315, MAG-RxDist-J11 to MAG-Extended Bridge Rear Support TNS Antenna, Phase-Length Controlled
5476790 Run M1335 MAG-AMP-J11 to MAG-RX Dist-J13, Cable, Body Receive, 5000 mm
5475852 Run M1352 MAG-BORE-J2 to MAG-DTRSW-J2, 3.0T Dual Body CH1, 3406 mm ref
5475852-2Run M1353 MAG-BORE-J3 to MAG-DTRSW-J3, 3.0T Dual Body CH2, 3406 mm ref
5498246-5Run M1354 MAG-DTRSW-J5 to MAG-AMP-J9, Body Receive, 710 mm
5498246-6Run M1355 MAG-DTRSW-J6 to MAG-AMP-J10, Body Receive, 710 mm
5475880 Run M1357 MAG-DPP1 J1 to MAG-Table IO-J17, DPP1 RF Signal 2130 mm
5475880-2Run M1358 MAG-DPP1 J2 to MAG-Table IO-J18, RF Signal, 2130 mm
5475880-3Run M1360 MAG-DPP2 J1 to MAG-Table IO-J20, RF Signal, 2570 mm
5475881-2Run M1359 MAG-DPP2-J3 to MAG-Table IO-J19, Pport BNC CONN 4, Coil Status, 2570 mm
5475880-4Run M1361 MAG-DPP2-J2 to MAG-Table IO-J21, RF Signal, 2570 mm
5475881-3Run M1362 MAG-DPP2-J4 to MAG-Table IO-J22, Pport BNC CONN 1Coil Status, 2570 mm
5475881-4Run M1363 MAG-DPP3-J3 to MAG-Table IO-J23, Pport BNC CONN 3, Coil Status, 3310 mm
5475880-5Run M1364 MAG-DPP3 J1 to MAG-Table IO-J24, RF Signal, 3310 mm
5475881 Run M1356 MAG-DPP1-J3 to MAG-Table IO-J16, Pport BNC CONN 2, Coil Status, 2130 mm
5475880-6Run M1365 MAG-DPP3-J2 to MAG-Table IO-J25, RF Signal, 3310 mm
5481324-2Run M1366 CABLE, Rx LO, DPP1-J10 to RxDistr-J7
5481324-3Run M1367 CABLE, Rx LO, DPP2-J10 to RxDistr-J6
5481324-4Run M1368 CABLE, Rx LO, DPP3-J10 to RxDistr-J5
5475870 Run M3337 MAG-SRI-J4 to MAG-LCP-J1, 4500 mm and MAG RCP J1, 1500 mm, Nonmagnetic (Compatible with 5475870-2)
5475870-2Run M3337, MAG-SRI-J4 to MAG-RCP-J1, 5,100 mm and MAG-LCP-J1, 2,150 mm
5495564 Run M3341 MAG-SRI-J20 to Table IO-J1,LANDMARK, D-sub Nonmagnetic, 3180 mm
5476792 Run M3342, MAG-SRI-J23 to Bore Patient Alignment Light
5788763Run M3344 MAG-SRI-J22, Bore Lights and Heat Exchanger Temp, Nonmagnetic (600 mm)
5549343 Run M3393, Rear Bore Light A (3200 mm and 2500 mm)
5549343-2Run M3394 Front Bore Light A (3200 mm and 2500 mm)
5475873-2Run M3345, MAG-SRI-J31 to MAG-Table IO-J3, Table Front Microphone, 3,970 mm
5476356 Run M3346, MAG-SRI-J32 to Table IO-J4, Rear Microphone, 3,950 mm
5183628 Run M3347 MAG RJ10 to MAG-Rear Endbell Speaker
5476794 Run M3348, MAG-SRI-J34 to RJ10, Comfort Module Speaker Cable, 4,600 mm
5475876 Run M3351, TCB-J5 to MAG-SRI-J2, Table Feedback, 3,400 mm
5476777-4Run M3367, MAG-DPP1-J7 to MAG-RxDist-J4, 44-Pin Sub-D, 1,150 mm
5476777-5Run M3368, MAG-DPP2-J7 to MAG-RxDist-J3, 44-Pin, Sub-D, 725 mm
5476777-6Run M3369, MAG-DPP3-J7 to MAG-RxDist-J2, 44-Pin, Sub-D, 350 mm
5475869 Run M3370, MAG-SRI-J3 to TCB-J6 Dock Control, 3,450 mm
5475860 Run M3380 MAG-Rx Dist-J58 to DTRSW J20, MAG-Bore-J9, MAG-PREAMP-J9, Hart ID Cable, Nonmagnetic
5476775 Run M3382 MAG-J61 to Parenthesis Light Left, 1000 mm
5476775-2Run M3383, MAG-J62 to Right Parenthesis Light, 4,000 mm
5476775-3Run M3506 MAG-Front-Upper Cover J63, 4000 mm
5495209-3Run M3386 Table IO-J6 to MAG-PAC-J5 J6, Peripheral Gating Cable for MR Kizuna, 3000 mm
5570806-2

See Note.

ECG Patient Lead Wire
Note:

5570806 is discontinued. Please order per following instruction.

  • If the color of housing is gray, order 5570806 until the existing stock is depleted. Then, order 5570806-2.

  • If the color of housing is white, order 5570806-2.

5476367-2ECG Cable Extension, mm
5495369 Run M3504, MAG-RxDist-J57 to TCB-J9 RxDist Power, 3,270 mm
5476753-4Run M4005 ISC GND STUD to MAG-GND STUD, 17500 mm
5476353 Run M3339, ICC PW-J118 to MAG-SRI-J33, Console Intercom, 15,600 mm
5475898 Run M3391 ISC-PW-J48 to MAG-Rx Dist J14, DPP ICE Diag,Nonmagnetic, 18500 mm
5476351 Run M3030 ICC PI J18 to E-OFF, 30000 mm
5492344 Run M3505 Extended Bridge-EOT to TCB-J10, D-Sub, EOT Switch, 5,900 mm
5561243 Run M3507 Extended Bridge-EOT Connect to Run M3505, EOT Switch, 200 mm
5476357 Run E3011 ISC-IO-J2 to OW-PortB, Ethernet, 30000 mm
5476626 Run E3018 OW-PDM-J3 to ISC-IO-J12, HART-RS422, 30000 mm
5756510USB Extender Cable FRU Kit, Improved Signal Quality (Including the following parts)
  • 5725077 Run E3037 USB Extender Cable, 30000mm

  • 5725077-2 USB Extender Cable, Cabinet Side, 2650mm

  • 1107-M5C010-37 Cross Recessed Pan Head Screw, M5-0.8 X 10mm Long

5479529 Run E3501 ISC-IO-J26 to Run E3047, 30000 mm
5502144 RUN E3042 ISC-IO-J112 to OW-PDM-J2, 30,000 mm
5476361-2Run E3014 ISC-PI-IO J20 to RF Door SW, E-Stop, 25000 mm
5476359 Run E0003, OW-PDM-J1 to ISC IO-J102, Host Power, 30,000 mm
5476361 Run E3030 ICC-IO-J115 to E-Off SW, E-off,17000 mm
5538322 GOC-PWM-J2 TO IEEE1394 and TO E3042
5476815 Run M1350 ISC-PW-J106 to MAG-DTRSW-J1, 3.0T Dual Body Tx,14463 mm ref
5476815-2Run M1351 ISC-PW-J109 TO MAG-DTRSW J4, 3.0T Dual Body Tx,14463 mm ref
5476647-4Run M3317 Gradient Cable X, Scan Room, Long, 12700 mm
5476647-5Run M3318 Gradient Cable Y, Scan Room, Long, 12700 mm
5476647-6Run M3319 Gradient Cable Z, Scan Room, Long,12700 mm
5475877 Run M1305 ISC- J122 TO Mag-Table IO J15, RF Transmit, 18379 mm
5476815-3Run M1350 ISC-PW-J106 TO MAG-DTRSW-J1, 3.0T Dual Body Tx, 10380 mm ref
5476815-4Run M1351 ISC-PW-J109 TO MAG-DTRSW J4, 3.0T Dual Body Tx, 10380 mm ref
5476647 Run M3317 Gradient Cable X, Scan Room, Short, 8700 mm
5476647-2Run M3318 Gradient Cable Y, Scan Room, Short, 8700 mm
5476647-3Run M3319 Gradient Cable Z, Scan Room, Short, 8700 mm
5475877-2Run M1305 ISC- J122 TO Mag-Table IO J15, RF Transmit, 14295 mm
6859057-2RUN E3040 ICC- J2 to ISC-J108, 4,500 mm
5475834-2RUN E0010 ICC_J5_ISC_J44, 5000 mm
5476823-2Run E0007 ISC-T2 to ICC-J1, 4000 mm
6859057 RUN E3040 ICC- J2 to ISC-J108, 10,000 mm
5475834 Run E0010, ICC-J5 to ISC-J44, 10,000 mm
5476823 Run E0007 ISC-T2 to ICC-J1, 10000 mm
5481324 Cable_TX481, Ref Clock, CLKS1-J3 to ICE PSC-J28
5481308 Cable, DVMRLINK, ICE-J29T to Exciter-J19
5481308-2Cable, DVMRLINK, PSC-J31T to GP4-J1
5486811 Cable, AC TRIGGER, ICE TO Cabinet Monitor, J43
5481312-2Cable_TX503, BD4, Driver Module-J25 ISC-PW-J96 Filter
5481312Cable_TX502, BD1, Driver Module-J21 ISC-PW-J95 Filter
5180907-53Run M3350 MAG-SRI-J10 to MAG-BORE, Bore Temperature, RoHS
Table 2. System Cable FRU for Remote Siting Configuration Only
Equipment Room Part#Description
5858947Run E3324 PW-J5 to ISC-IO-J143, DCPS for IRD
5858948Run E3314, PW-J42 to ISC-IO-J42, Display Power
5858952Run E3311 ISC-IO-J36 to PW-J36, D-sub, Long Drive Power
5477605-15Run P2004 ISC-ICE-J6, 13, 14 to MAG-DPP2 J9, DPP3 J6, 9, 34.7m
5476646-6Run P2011 MAG-SRI J13 J14 to ISC-PHPS2-J37_J38, CAN Fiber Optic, 34.15m
5476646-7Run P2010 MAG-SRI J11_J12 to ISC-ICE-J40_J41, CAN Fiber Optic, 34.4m
5861075Remote Siting (Off the Wall) Cable Run E3300 ISC-IO-J112 to PW-J2, Rx Dist Power, 24W7, Low-Mag
5858971Run E1302 ISC-PW-J29 TO ISC-IO-J29, Receiver Clock, Twinax, 17.50m
5860964Remote Siting (Off the Wall) Cable Run E4005 ISC-GND to PP-GND
5881836Remote Siting (Off the Wall) Gradient Cable FRU, 17m, Equipment Room Side
5881837Remote Siting (Off the Wall) Gradient Cable FRU, 13.5m, Scan Room Side
5747058Run E1051, ISC-J109 to PEN-J109, 3.0T Body, 10,327 mm
5747059Run E1001, ISC-J106 to PEN-J106, 3.0T Body, 10,327 mm
5747768Run E1002, ISC-J122 to PEN-J122, RF Transmit, 10,218 mm
5747058-3Run M1351, DTRSW-J4 to PEN-J109, 3.0T Body, 10,500 mm
5747059-3Run M1350, DTRSW-J1 to PEN-J106, 3.0T Body, 10,500 mm
5747761Run M1305, PEN-J122 to Mag-HDTX-IO-J2, P2 RF Transmit, 14,305 mm
5747768-2Run E1002, ISC-J122 to PEN-J122, RF Transmit, 16,348 mm
5747058-2Run E1051, ISC-J109 to PEN-J109, 3.0T Body, 16,523 mm
5747059-2Run E1001, ISC-J106 to PEN-J106, 3.0T Body, 16,523 mm
5747058-4Run M1351, DTRSW-J4 to PEN-J109, 3.0T Body, 14,631 mm
5747059-4Run M1350, DTRSW-J1 to PEN-J106, 3.0T Body, 14,631 mm
5747761-2Run M1305, PEN-J122 to MAG-HDTX-IO-J2, P2 RF Transmit, 18,392 mm
Table 3. Run P5006 Cable Harness, GOC to IRD Fiber Optic-5875244 for MR Compatible 12.1 inch In Room Display-8771126-61
Part NumberDescription
5778287-8USB-AOC optic fiber cable, Host PC USB3.0 port to IRD-Right-J3, Run# P2027 48000mm, GEN2
5778287-9USB-AOC optic fiber cable, Host PC USB3.0 port to IRD-Left-J3, Run# P2029 48000mm, GEN2
5879076DP AOC, GOC DP Dongle to IRD Right DP, Run P2026 48000mm
5879076-2DP AOC, GOC DP Dongle to IRD Left DP, Run P2028 48000mm
5851077-2 Mini DP to DP Dongle for port 1
5851077-3Mini DP to DP Dongle for port 2