LVShim Process

LVshim provides an accurate method of collecting and analyzing scan data to determine and correct magnetic field inhomogeneity. This procedure has been updated to include the TwinSpeed (TRM) which has Whole-Body (WB) and Zoom (ZM) gradient coils. The calibration only needs to be done on the Whole-Body coil. LVshim consists of four procedures: Rough, Main, Gradient and Passive Shim. It is important when performing LVShim that all prerequisites are completed and that the proper order is followed.

note:

When doing LV shim as the result of an upgrade, the recommendation is to re-enter the last set of shim currents into the magnet and start with Main LV shim. You should not have to do Rough LVShim if the initial system had a reasonable shim prior to the upgrade.

note:

After ramping the magnet, the cryocooler coldhead must be turned off for 10 minutes before installing shim currents. This is due to the magnetic flux buildup within a superconducting shield around the coldhead sleeve during magnet ramping.