LVShim Software
The LVShim software consists of three elements:
-
The LVShim PSD: Defines the pulse sequence data base (PSD) used to acquire the LVShim scan data.
-
LVShim Calibration File Generator: Generates LVShim calibration files.
-
LVShim Analysis Tool: Analyzes the scan data acquired by the LVShim PSD, and computes the magnetic field inhomogeneity and new shim current settings.
The subjects covered in this document:
-
Magnet Serial Number, Magnet Serial Number
-
Calibration File Type, Calibration File Type
-
Calibration File Name, Calibration File Name
-
Shim Type, Shim Type
-
Image Data. Image Data
-
Operation Mode, Operation Mode
-
Display Magnitude and Phase Difference Images, Display Mag and Phase Diff Images
1 Overview
LVShim calibration files contain coil characterization data used by the LVShim analysis tool to calculate the new currents needed to correct for magnetic field inhomogeneity. Generic shim calibration files are supplied with the system software for all gradient coils and for GE S/C shim coils.
2 Magnet Serial Number
The magnet serial number is used by LVShim to determine the appropriate set of calibration files to display during the shimming process. The magnet serial number must be correctly entered in the MR Configuration file. (The Magnet Type field is no longer used by system software tools.) Table 1 shows the value for each magnet type.
3 Calibration File Type
The software displays only the Calibration file types appropriate for the type of magnet being shimmed. Refer to Table 2.
4 Calibration File Name
Figure 1 is an example of a shim calibration file with the name components explained.
Figure 1. Breakdown of a Shim Calibration File Name

Figure 2. Example of Gradient Shim Calibration File

Refer to Table 3 for a list of the available coils.
The S/C shim calibration files are divided into magnet/coil type, then by four file types: 22dsv.full, 32dsv.full, 40dsv.full, 45dsv.full, 40dsv.half, and 45dsv.half. The number signifies the diameter of the spherical volume.
-
The 22dsv.full files are used during Rough and Gradient LVShim.
-
The 32dsv.full files are for troubleshooting purposes only.
-
The 40dsv.full (for CRM Body Coil) and 45dsv.full files are used during Main LVShim.
-
The 40dsv.half and 45dsv.half files are half-current files used in cases where the full file is "overshooting" the target value.
Table 4 lists the generic calibration files supplied with the system.
5 LVShim Analysis Tool Details
5.1 Shim Type
The software displays only the shim types associated with the magnet being shimmed. Table 5 shows the shim types for each magnet.
When Test is selected as the Shim Type, new shim currents are not calculated during LVShim analysis.
5.2 Image Data
Image Data refers to the exam, series, and image number of the image data set to be processed. If an LVShim scan was just performed, the program automatically picks up the last exam, last series, and first image number of the last set of images scanned and makes it the default choice. If the last scan was not an LVShim scan, Image Data is shown as none selected.
Select this menu item when the image data displayed by the software are incorrect. When Image Data is selected, the software prompts for the correct exam, series, and image numbers. If the exam, series, or image number of the image data to be processed is not known, enter 0 (zero) when prompted for a number; this provides a list of the existing exams, series, or images.
For the TwinSpeed, take care to select the series corresponding to the same GradMode highlighted upon entering the LVShim Analysis Tool.
5.3 Operation Mode
There are two operation mode options: Service and Research, and Table 6 describes them.
5.4 Display Magnitude and Phase Difference Images
This option allows the display of each magnitude image and masked phase difference image in a separate window (shown in Figure 3) after the image data is processed. Both windows are updated when an alternate scan plane is selected. These windows will initially display in the upper right corner of the screen; reposition as necessary for viewing the LVShim Analysis window.
Figure 3. Magnitude and Phase Difference Images

5.5 Calibration File Name Token
The calibration file name token is used to help distinguish similar calibration file names from each other. The token is attached at the end of the calibration file name. Refer to Calibration File Name, Calibration File Name, for more information.