Grafidy 3 Gradient Calibration

Grafidy 3 utilizes Grafidy PSD (pulse sequence) and Grafidy coil fixture, and has an improved user interface to allow more flexibility and ease of use. Grafidy 3 will adjust and compensate for short, long and very long time constants eddy currents, both linear and B0 terms. It is expected that the FE will start the calibration process by running Grafidy 3 in Auto Mode and then switch to Manual Mode only if Grafidy 3 cannot bring a term into specification. This should not often have to be done.

Refer to Table 1 for the suggested Grafidy 3 flow.

Additional information can be found for GE Service Pull-Down and Aborting/Exiting the Grafidy 3 tool.