XRMw Shim Retaining Ring Segment Replacement

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-360 minutes-
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
XRMw gradient coil shim retaining ring FRU kit (contains 6 screws and pack of sandpaper)1-

5431091

-
Non-ferrous toolkit1-

5112581

-
B0 quench box1-

2141701

-
Table 3. Consumables
ItemQuantityEffectivityPart numberManufacturer
SandpaperAs needed---
Table 4. Safety
Table 5. Required conditions
ConditionReferenceEffectivity

Remove patient front end bell and patient support bridge from the magnet bore as described in appropriate procedures before starting work.

--

About this task

Overview

This procedure describes how to replace the shim retaining ring on the XRMw gradient coil while the magnet is ramped. The retaining ring is split into 3 segments, which can be replaced individually. The segments are split between shim trays 36 and 0, trays 12 and 13, and trays 24 and 25.

Figure 1. Shim Retaining Ring

Procedure

  1. Remove the passive shim trays from the gradient coil in the section where the ring segment is to be replaced.

    Refer to the Magnet and Cryogens Subsystem Manual for the shim tray removal process.

  2. Using non-ferrous Phillips screwdriver and non-ferrous channel lock pliers, remove the 6 stainless mounting screws on the segment of the ring to be replaced.

    You may need to tap on the screw heads to break the Loctite bond before you are able to remove them.

    Figure 2. Removing Stainless Mounting Screws
  3. Pry loose and take out the broken ring segment. Start prying from the end of the segment to help break the epoxy bond.
  4. Using sandpaper, clean the mounting surfaces on the gradient coil of any resin fragments to ensure a good fit of the ring segment (especially the mounting surface on the inner surface of the outer gradient coil).
  5. Replace the broken ring segment with the new one.
    Figure 3. XRMw Gradient Coil Shim Retaining Ring
  6. Safely re-install all passive shim trays. Be sure to return each tray to the correct slot to avoid changing the shim of the magnet.

Finalization

  1. Re-install the bridge and front end bell.
  2. Connect the B0 coil power supply output cable to P304-1 on the magnet connector box.
  3. Connect the B0 coil power supply input cable between the power supply and the voltage source.
  4. Press the B0 coil power supply switch (the switch is spring-loaded) and hold for one minute.

    The LED lights to indicate the power supply is supplying output current.

  5. Wait for three minutes for the magnet to recover before performing LVShim.
  6. Perform an LVShim scan.
  7. Perform a Check Scan.