- 00000018WIA301E4E20GYZ
- id_131064863.7
- Jan 11, 2021 12:20:17 PM
HEC Hose Inspection, Coolant Fill, and Coolant Leak Check
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 90 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| 5 gallon pail | 1 | - | - |
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Coolant | As required (48 one-gallon containers shipped with the system) |
5174313 | - |
| Paper towels (white) | 1 roll | - | - |
| Teflon (PTFE) tape | As required | - | - |
| Required conditions |
|---|
| Shutoff (ball) valves are open. |
| Hose connections are secure. |
| Check valves are correctly oriented. |
| Safety |
|---|
| Warning | |
|---|---|
| Notice | |
|---|---|
About this task
- Inspect the hoses and valves to ensure that they are correctly installed.
- Fill the gradient coil and power electronics tanks in the Heat Exchange Cabinet (HEC) with a deionized water solution. (Usually, the gradient coil tank is filled before the power electronics tank.)
- Check the valves for leaks after the pumps are activated and the fluid is circulating.
- Check the joints are located outside of cabinet perimeters.
Note: If you are doing Planned Maintenance, please execute this procedure completely when you see significant coolant loss (more than a gallon) or any pump leak issues. Otherwise perform only the Filling the coolant procedure to top off the coolant tanks.