• Optima MR450w BASE 1.5T System Service Methods
  • 5690012-2EN Revision 3
  • Object ID: 00000018WHA30A99BGYZ
  • Topic ID: id_2002114 Version: 8.3
  • Date: Jul 9, 2020 11:22:46 AM

Running Grafidy

Running Grafidy in auto mode includes all axes.

Procedure

  1. Start the Grafidy tool.
    • (For non-proprietary service tools) From the Common Service Desktop, select Calibration > Grafidy. Select Click here to start this tool.

    Result

    The Grafidy tool window appears.
  2. Determine if the existing terms should be zeroed out.
    • For new system calibrations, click on the Zero Out tab.

      Select all the terms on the left and click Zero Out Cal Values before proceeding to the next step.

    • For performance checks only, proceed to the next step.
  3. Select the Auto Mode tab. All three coil axes are selected automatically.
  4. Click Scan to start the calibration.
    • When the scan is active, it can be stopped by clicking Abort on the window and pressing Stop on the keyboard.
    • Highlight the desired axis and click Show to display the result graphs.

    In auto and manual modes, a number of signal checks are performed to ensure the proper state of the Grafidy fixture. The following signals are examined from every channel. An error is reported if a problem is found.

    • Position
    • Signal-to-Noise Ratio (SNR)
    • Signal saturation

    The Grafidy software runs multiple scans to bring each term into specification. It iterates up to the maximum amount specified (default is five iterations). When each component finishes a scan/analysis iteration, the residual of the eddy current of the measurement displays in the results table. The results are color-coded.

    Text colorDefinition
    RedValues above specification
    GreenValues within specification

    Auto mode acquires data in the following order:

    • DC
    • Vlong
    • Long
    • Short
    • Long (x & y) - second pass
  5. Review the numeric output values from the auto mode scan. Verify that the results of the final iteration for each component are within specification (green).
    Figure 1. Reading output with specifications and actual values (smoothed)
  6. If a component is not within specification after the maximum number of iterations, use manual mode to calibrate that component.