- Optima MR450w BASE 1.5T System Service Methods
- 5690012-2EN Revision 3
- 00000018WIA30722750GYZ
- id_20255731.2
- Nov 5, 2020 1:37:09 PM
Probe tuning and SNR check troubleshooting
About this task
This section contains illustrations of probe SNR spectra, both incorrectly and correctly tuned, to show what a good spectrum looks like. During tuning or SNR troubleshooting, it may be necessary for a site to back up the original echoloc.dat file. This backup process requires renaming the file; otherwise, it will be overwritten during the tuning calibration. To get more information about the cause of an Auto-Water Suppression (AWS) failure, log the messages generated during the Probe APS. A site may be required to send probe SNR raw data to an expert for review. This section also explains the P raw file location process and how to transfer it.
Example SNR Spectra
About this task
Errors in the probe tuning calibration procedure can introduce changes in the appearance of the resulting probe SNR spectra. This section contains example spectra from a Probe-P SNR protocol using the MRS phantom.
Procedure
Directory Creation
Procedure
P Raw Data File Location
About this task
During the troubleshooting process, the site may be required to identify the P raw data files generated when Probe SNR was performed. These files may be helpful in determining a reason for an SNR failure. This portion of the procedure should be performed after running a Probe SNR protocol to facilitate P raw file identification.