Synchronizing NTP server time
Network Time Protocol (NTP) server configuration can be done to synchronize the MR scanner date and time automatically with an NTP server available on-site. An NTP server and IP address are required.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 30 minutes | - |
| Required conditions |
|---|
| The procedure in this document requires that the site has an NTP-based time server. During the procedure, you will need to enter the IP address of the NTP server. Before you begin, get this address from the site's network administrator or IT department personnel. |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
Important: Make sure the MR system is not scanning before starting this procedure.