Installing Service Packs
When GE releases new software versions, it may be necessary to install updates to a host PC.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 30 minutes | 10 minutes |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Service Key | 1 | ||
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| GE Equipment Maintenance Sticker | 1 | 5661793 | N/A |
| Required conditions |
|---|
| Condition |
| Updates are loaded after MR apps and service software have been loaded. Service Packs must be loaded at root level (applications not up). |
| Safety |
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
Important: Make sure that the latest Service Pack is loaded. See the latest revision of the MR Software Release Matrix, DOC1667089, available from the online documentation library.
Note: The Remote Software Download application can manage software updates and install them automatically. Refer to Installing updates from Remote Software Download (RSD) for more information.
Important: Open the “Read Me” file to identify any special procedures that need to be completed in order to install the service pack.