• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WHA3062A4GYZ
  • Topic ID: id_2001781 Version: 16.1
  • Date: Jan 25, 2022 10:38:18 AM

Applying LOTO - ISC

Apply LOTO to both the gradient and system PDU breakers on the MDP when servicing either of these items in the ISC. Applying LOTO ensures a safe environment when working on these parts.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-
Procedure details
Name of equipment Number of locks Titles of employees authorized to perform LOTO Titles of affected employees How to notify
  • ISC
  • Electronics attached on magnet (except magnet)
  • Patient table
  • GOC
1 per GE field engineer (FE) GE FEsHospital personnel Verbally, posted signs
Tools and test equipment
ItemQuantityPart numberManufacturer
Digital Voltmeter (DVM)146-194427P284-
Red Warning LOTO Tag246-194427P322-
Multi-Locking Device (if multiple service personnel are involved)146-194427P313-
Red LOTO Personal Lock Wrap22393068-
Brass Master Padlock with Brass Shackle246-194427P320-
Note: The following table describes the type, location, and magnitude of energy when applying LOTO.
Energy sourceLocation of energy isolating meansMagnitude of energy
ElectricalMDP380/400/415/480 VAC
Stored energyTime discharge of internal capacitors208 VAC, 420 VAC
Water under pressureICC5.0 Bar (80 PSI)
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

Make sure the host computer is shut down.