• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA30E48030GYZ
  • Topic ID: id_12374319 Version: 1.1
  • Date: Jul 5, 2019 10:03:33 PM

Manual Entry Probe/SV SNR Protocols

This document contains the Non-Proprietary protocols for Testing Probe-S and Probe-P SNR. Refer to Probe/SV Calibration and SNR Tests for the complete calibration procedure, and substitute the protocols below for the Service protocols listed.

  1. Refer to Table 1 for the non-proprietary Probe-S SNR Protocol.

  2. Refer to Table 2 for the non-proprietary Probe-P SNR Protocol.

Table 1. PROBE-S SNR PROTOCOL

Patient Register New Pt

PATIENT INFORMATION

Patient ID geservice

Patient Name probes snr

Weight (Lb) 111 (IMPORTANT) Landmark

Landmark >Nasion

Patient Protocols Patient Position

Patient Position

Patient Position >Supine

Patient Entry >Head First

Coil ...HeadAccept

Imaging Parameters

Plane >Axial

Mode >MRS

Pulse Seq ...Probe-SAccept

Imaging Options Ext. Dyn. Range

Psd Name (no entry)

Protocol (no entry)

Scan Timing

# of Echoes 1 (default)

TE 30

TR 2000

User CVs Screen

CV3 scan mode1.00

CV4 total # of scans32.00

Scanning Range

FOV24

Start S10

EndI10

StartR10

EndL10

StartA10

EndP10

Table Delta0.00 (default)

Acquisition Timing

Freq1

Phase1

NEX2.00

Freq Dir>A/P

Auto Center Freq>Water, Autoshimselected, Contrastnot selected

Save series

Prepare to Scan

Research Operations Display CVs CV Name CV Value

tempC (greenest value on MRS Phantom) Accept

Research Operations Download

Table 2. PROBE-P SNR PROTOCOL

Patient Register New Pt

PATIENT INFORMATION

Patient ID geservice

Patient Name probep snr

Weight (Lb) 111 (IMPORTANT) Landmark

Landmark>Nasion

Patient Protocols Patient Position

Patient Position

Patient Position>Supine

Patient Entry>Head First

Coil...HeadAccept

Imaging Parameters

Plane>Axial

Mode>MRS

Pulse Seq...Probe-PAccept

Imaging Options Ext. Dyn. Range

Psd Name (no entry)

Protocol (no entry)

Scan Timing

# of Echoes 1 (default)

TE 35

TR 2000

User CVs Screen

CV3 scan mode 1.00

CV4 total # of scans 32.00

CV15 Phantom = tempC value

CV17 = 0

CV18 = 7

Scanning Range

# CSI Slices1

FOV24

Start S10

EndI10

StartR10

EndL10

StartA10

EndP10

Table Delta0.00 (default)

Acquisition Timing

Freq1

Phase1

NEX2.00

Freq Dir>A/P

Auto Center Freq>Water, Autoshimselected, Contrastnot selected

Save series

Prepare to Scan

Research Operations Display CVs CV Name CV Value

tempC (greenest value on MRS Phantom) Accept

Research Operations Download