Leak Sensor 2 (Front)
Prerequisites
Procedure
- Remove L Upper Front Cover. Refer to SC Cover Removal
- Disconnect the two connectors of leak sensor 2 (Front).
- Remove the leak sensor 2 (Front) by loosening 6 screws.
Figure 1. Leak Sensor (Front)

- Installation is reverse order of removal.
1 Finalization
Procedure
- Restore the Power. Refer to System Cabinet PDU Main Breaker LOTO Procedure.
- Perform Cabinet Monitor Functional Check.
- Run one head or body scan.