Table Top Height Adjustment
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 10 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Non-ferrous standard head screwdriver | 1 | - | - |
Overview
This procedure provides instructions on how to adjust the height of the table top.
Procedure
- Dock the patient table.
- Use the UP foot pedal to raise the patient table to full up
position, and move the cradle completely into the magnet bore to expose
the table top height adjustment screw.
Figure 1. Table top height adjustment screw location

1 Table top height adjustment screw 2 Front of table - Use a standard head screw driver to remove the cover plug.
- Adjust the screw to set the table top height equal to the bridge
height.
Figure 2. Rolling surface on bridge and table

Item Description 1 Cradle 2 Rolling surface (path) on left and right sides of bridge and table top. Straight edges are used to show the paths. 3 Bridge 4 Table - Replace the cover plug.
Finalization
- Dock the table.
- Confirm that the cradle release functions properly.
- Confirm that the LPCA connects to the cradle and that the cradle can move in and out of the bore.