Running dual drive quadrature tool

Procedure

  1. Click Start for the first iteration.

    Scan and analysis begin. It takes about 6 minutes.

  2. The tool runs up to three iterations, attempting to come to specification.

    When the first iteration is complete, calibration results are passed or failed.

  3. (For passed) Click Quit to complete the calibration. Proceed to Finalization.

    Figure 1. Pass case

  4. If the calibration passes within iteration 3, click Quit to exit the tool.