1.5T Neurovascular Array Setup for SNR Test
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 15 minutes | Not Applicable |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Phantom and positioner kit | 1 | - |
5117092-6 |
- |
| Condition | Reference | Effectivity |
|---|
Follow this process to prepare for the automated SNR test using the 1.5T NV Array by Medrad (M3087JJ).
Procedure
- Position the coil on the table.
- Place the positioner and phantom in the coil. Do not use any pads when running the SNR test.
- Press the chest section down till resting on phantom. See Figure 1.
Figure 1. Coil Loaded with Phantoms for SNR Test

- Slide the head section of the coil all the way in the inferior direction, till it locks in place. See Figure 1.
- Landmark on the coil crosshairs on the top part of the head section. See Figure 1.
- Perform the MCQA Test if available, otherwise perform the Surface Coils SNR Test.
Finalization
No finalization steps.