- Topic ID: id_17479539
- Version: 2.0
- Date: Nov 19, 2019 11:08:55 PM
Primary Aperture Replacement
Prerequisites
Overview
This procedure provides the following:
-
Remove gantry covers and power down.
-
Remove X-Axis Tube Gauge Mount Bracket and Collimator Assembly, Interposer Plate, and primary aperture.
-
Install new primary aperture and reassemble components in reverse order of disassembly.
-
Finalization procedures.
|
note:
Please check the component P/N when replace the collimator related parts, should be ordered the same P/N according to the label on each component.
Procedure
- Remove gantry power. Be certain to follow appropriate Lockout/Tagout procedures.
- Remove the gantry side, top and front cover.
- Turn OFF all 3 switches (AXIAL DRIVE ENABLE, HVDC ENABLE, 120 VAC) on the Service Switch Panel.
- Position the gantry with the X-Ray Tube at six-o'clock.
- Remove the four hexagon socket M4 screws for the X-Axis Tube Gauge Mount Bracket.
- Remove the X-Axis Tube Gauge Mount Bracket.
- Follow the procedure for removal of the Collimator Assembly.
- Remove the two slotted head shoulder screws for the Interposer Plate.
- Remove the Interposer Plate with Interposer Adjuster Block attached.
- Remove eight countersunk flat head screws to remove the primary aperture.
- Remove the aperture and place in the container for shipment.
- Install the new primary aperture ensuring not to damage any of the interior edges due to the critical nature of x-ray filtering.
- Apply Loctite 242 to the eight flat head screws and tighten.
- Apply Loctite 242 to the two shoulder screws and mount the Interposer Plate tightening the screws.
- Apply Loctite 242 to the four socket screws and mount the X-Axis Tube Gauge Mount Bracket tightening the screws.
- Follow the procedure for mounting the Collimator on the gantry.
Finalization
- Perform Plane of Rotation (POR).
- Perform Beam on Window Alignment (BOW).
- Perform CBF/SAG Alignment Process.
- Perform ISO Alignment Procedure.
- Perform Detailed Phantom calibrations.
- Perform Image Series Check.