• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30C89230GYZ
  • Topic ID: id_15667346 Version: 1.6
  • Date: Jul 13, 2021 4:27:37 PM

Touch-n-Go (TnG) switch replacement

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable60 minutes45 minutes
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Non Magnetic Tool Set1---
Table 3. Replacement parts
ItemQuantityEffectivityPart numberManufacturer
TnG Switch 1-

Refer to FRU Manual

-
Table 4. Safety

Procedure

  1. Perform the following steps to remove 2 covers.
    1. Remove 5 screws and slide out the Motor Cover.
    2. Remove two rubber caps, remove 6 screws, and remove side cover.
    Figure 1. Cover removal
  2. Perform the following steps.
    1. Remove 11 nuts.
    2. Remove a connector.
    3. Take out TnG Switch Assy.
    Figure 2. TnG removal
  3. Align the location of TnG Switch Assy and restore it per following steps.
    1. Insert 11 studs of TnG Switch Assy into the holes.
    2. Loosely screw the 11 nuts to the 11 studs.
    3. Adjust the TnG Switch Assy to make the TnG surface aligned with Table top FRP cover.
    4. Tighten the 11 nuts.
    5. Observe if the TnG surface is aligned with table top FRP cover or not. If not aligned, perform Step 3.b through Step 3.d until the surface is aligned, then execute next step.
    6. Connect the TnG cable connector.
    Figure 3. TnG switch assembly location alignment
  4. Restore Patient Table by reverse order of removal.

Finalization

  1. Turn the system power ON. Refer to Lockout / Tagout for ISC.
  2. Check Touch and Go function. Refer to Table Functional Check .
  3. Perform one test scan.