• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30882110GYZ
  • Topic ID: id_2003539 Version: 3.3
  • Date: Jul 7, 2021 3:51:55 PM

Removing the magnet left-hand rear cover

Removes the rear cover from the left-hand side of the magnet.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-75 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

Do the following tasks to complete removing the magnet left hand rear cover:
  • Removing the magnet rear trim ring
  • Removing the magnet rear cable cover
  • Removing the rear bridge cover
  • Removing the rear bridge inner cover
  • Removing the rear endbell
  • Removing the magnet left-hand rear cover