• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WHA30B00CGYZ
  • Topic ID: id_2002356 Version: 6.1
  • Date: Feb 7, 2022 4:45:45 PM

Loading software updates from repository or CD/DVD

You can install software updates to the host computer from the repository or from a CD/DVD.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-
Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Class M service key1--
Service Pack CD/DVD1-Not applicable
Required conditions
Condition
Load updates after MR apps and service software have been loaded.
Service Packs must be loaded at root level (applications not up).
Invoke Guided Install as root and display the Patches tab. See Invoking Guided Install as root.

About this task

Note: Updates should not be loaded with application software running.

Procedure

  1. Either insert the update CD/DVD in the appropriate drive and select CD/DVD, or select Repository to load updates from /usr/g/insite/data.
    Figure 1. Guided Install patches
  2. Click the applicable update to load and select Install.
    Note: You must load one update at a time.
    Figure 2. Available updates

    Result

    Updates that load successfully show in the Installed Patches window. A confirmation message also appears in the GUI status section.
    Figure 3. Install done message

    The update process looks for the software revision and conflicting information before loading. If the software revision of the update does not match the applications revision, a message shows in the status section of the GUI that explains that the update cannot be loaded.

    If the message Patch_tst_M4_01 = PA conflicts with patch_tst_M4_03-PA-1 appears, you must delete the most recently installed update from the system before you load the new update. See Removing software updates.

  3. Do any other patch load instructions that are documented in the “Read Me” file of the Service Pack.
  4. Restart the system. When prompted, log on as sdc for user.

Finalization

Do a check scan to make sure that the system is functioning properly. See Doing a check scan.