• Discovery MR750w and SIGNA™ Architect T 3.0T System Service Methods
  • 5690002-2EN Revision 4
  • Object ID: 00000018WIA30F7EE20GYZ
  • Topic ID: id_13106742 Version: 3.0
  • Date: Aug 29, 2019 1:46:29 AM

Respiratory Gating Signal Checks

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable25 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Respiratory Bellows1-

46 265679G1

-
Table 3. Safety

Procedure

  1. Connect bellows to the Enclosure Interface module.
  2. Go to Tools tab.
  3. Select Gating Control menu.
  4. Select Respiratory. The Respiratory Waveform appears
  5. Take the respiratory bellows in both hands and gently pull on either end.
  6. Pull the bellows by hand duplicating the motion of typical regular breathing. If the bellows and respiratory compensation software is working properly the LED's should follow the actions of the bellows. If it does not, proceed by troubleshooting the bellows (for tears or holes), connections and then the PAC assembly for problems.
  7. Verify waveform is displayed following action of bellows.
    Note:

    The respiratory waveform will be displayed on the IRD if the system is so equipped. If not, it will be displayed on the Enclosure display.

Finalization

No finalization steps.