• Discovery MR750w and SIGNA™ Architect T 3.0T System Service Methods
  • 5690002-2EN Revision 4
  • Object ID: 00000018WIA30CE9E20GYZ
  • Topic ID: id_13106478 Version: 2.2
  • Date: Jun 22, 2020 3:16:33 PM

Guided Install Site Configuration Options

The tables below describe some options available when configuring a site system using Guided Install. Only tabs containing system-dependent information are described below.

To see additional information about a field in the Guided Install interface, put the cursor in that field and look at the bottom line of the screen. For additional information about a tab, use the Help menu.

Note:

For Artist XT and Architect XT systems, option keys must be installed for system handover.

Table 1. System Configure
Parameter 450 450w/Artist T Artist/Artist XT 750 750w/Architect T Architect/Architect XT
Hospital Name
Suite_ID
Host_ID
Unique/System ID
Service ID
Date Format
Weight Unit
Keyboard
Language
Video CardN/AN/AN/AN/AN/AN/A
Display Type

Model type of the monitor.

For the NEC P241W monitor: Select LCD Monitor. If this choice is not available, select Other.

Enable HO ShimSet to Yes if HO Shim is included on the systemSet to No (High Order Shim is not available)Set to No (High Order Shim is not available)Set to Yes if HO Shim is included on the systemSet to No (High Order Shim is not available)Set to No (High Order Shim is not available)
Legacy Image Type

Select Yes to share images with a workstation that does not support DICOM.

Scan Range (micro-meter)

If the length of the magnet room is less than 21.5 ft (6.5 m), select Short. If the length is greater than 21.5 ft (6.5 m), select Long.

(For DV26.0 and later) This field is located on the hardware configure tab.

Long or Short Long or Short Long or Short Long or Short Long or Short Long or Short
Table 2. Hardware Configure (prior to DV26)
Parameter 450 450w 750 750w
Field Strength1.5T1.5T3.0T3.0T
Package Name

SR (120 or 200) is automatically updated based on XRM or XRMw. Focus only on XGD or XG2 for the Package Name selection. Select XGD SR120/SR200 if the PGR cabinet has a Cronus Chassis, or select XG2 SR120/SR200 if it does not have a Cronus Chassis.

XGD SR200 or XG2 SR200XGD SR120 or XG2 SR120XGD SR200 or XG2 SR200XGD SR120 or XG2 SR120
Gradient Type

Select 8920 (XGD) if the PGR cabinet has a Cronus Chassis, or select 8921 (XG2) if the PGR cabinet does not have a Cronus Chassis.

8920 (XGD) or 8921 (XG2)8920 (XGD) or 8921 (XG2)8920 (XGD) or 8921 (XG2)8920 (XGD) or 8921 (XG2)
Resonance ModuleXRMXRMWXRMXRMW
Magnet Serial Number

Enter the actual magnet serial letters and numbers, found on the magnet rating plate on the rear of the magnet. This entry is case sensitive.

Magnet Ramp DirectionForwardForwardForward Forward
Table LimitAuto-populatedAuto-populatedAuto-populatedAuto-populated
Line Frequency 50 or 60 Hz 50 or 60 Hz 50 or 60 Hz 50 or 60 Hz
RF Amp Type1.5T SRFD2 RF Amp or 1.5T 16KW XRFD1.5T SRFD2 RF Amp, 1.5T 16KW XRFD, or Dual Drive3.0T 35kW XRFD or 3.0T 30kW Dual Drive RF AmpDual Drive
Governing Body IEC IEC IEC IEC
Iso Vector Z1.5T 60 cm Magnet Enclosure1.5T 70 cm Magnet Enclosure3.0T 60 cm Magnet Enclosure3.0T 70 cm Magnet Enclosure
Magnet EnclosureDV EnclosureDV Enclosure or GEM EnclosureDV EnclosureGEM Enclosure
PAC TypePAC2B/PAC5PAC2B/PAC5PAC2B/PAC5PAC2B/PAC5
RF-Systems Cabinet (RFS)YesYesYesYes
Spectro RF Amp TypeNone or CPC 5118154 or CPC 5118154-3 None if site does not have spectroscopy option, or AMT if site has spectroscopy option hardware None if site does not have spectroscopy option, or AMT if site has spectroscopy option hardware None if site does not have spectroscopy option, or AMT if site has spectroscopy option hardware
Scanner Channel Configuration16 or 3216 or 3216 or 3232
Shim Supply TypeNAV Shim Supply if HO Shim is presentNoneNAV Shim Supply if HO Shim is presentNone
LPCA Configuration Type2 P-ports on LPCA1 P-port on LPCA or 2 P-ports on LPCA2 P-ports on LPCA1 P-port on LPCA
HEC Type

Click More info to see details.

G6000EN or G6001EN G6000EN or G6001EN G6000EN or G6001EN G6000EN or G6001EN
UPM TypeUPM 100KHz sampling rate or QUPM 1 MHz sampling rateUPM 100KHz sampling rate or QUPM 1 MHz sampling rateUPM 100KHz sampling rate or QUPM 1 MHz sampling rateUPM 100KHz sampling rate or QUPM 1 MHz sampling rate
Table ConfigurationNoneNoneNone Flat (GEM) table if the site has at least one flat table; Curved table if the site has ONLY curved tables
Table 3. Hardware Configure (For DV26 and Later)
Product Name 1.5T, 60cm, Discovery MR4501.5T, 70cm, Optima MR450w1.5T, 70cm, Optima MR450w GEM1.5T, 70cm, SIGNA Artist T1.5T, 70cm, SIGNA Artist1.5T, 70cm, SIGNA Artist XT3.0T, 60cm, Discovery MR7503.0T,70cm,Discovery MR750w 3.0T, 70cm, SIGNA Architect T3.0T, 70cm, SIGNA Architect3.0T, 70cm, SIGNA Architect XT
Field Strength 1.5T1.5T1.5T1.5T1.5T1.5T3.0T3.0T3.0T3.0T3.0T
Package Name XGD

XG2

XGD

XG2

XGD

XG2

XG2XG2XG2XGD

XG2

XGD

XG2

XG2XG2XG2
Gradient Type 8920(XGD)

8921(XG2)

8920(XGD)

8921(XG2)

8920(XGD)

8921(XG2)

8921(XG2)8921(XG2)8921(XG2)8920(XGD)

8921(XG2)

8920(XGD)

8921(XG2)

8921(XG2)8921(XG2)8921(XG2)
Resonance Module XRMXRMWXRMWXRMWXRMWXRMWXRMXRMWXRMWXRMWXRMW
Magnet Serial Number RHMHMHMHMHMWUAUAUAUA
Magnet Ramp Direction ForwardForwardForwardForwardForwardForwardForwardForwardForwardForwardForward
Table Limit
Scan Range Long

Short

Long

Short

Long

Short

Long

Short

Long

Short

Long

Short

Long

Short

Long

Short

Long

Short

Long

Short

Long

Short

Line Frequency 50Hz

60Hz

50Hz

60Hz

50Hz

60Hz

60Hz

50Hz

60Hz

50Hz

60Hz

50Hz

50Hz

60Hz

50Hz

60Hz

60Hz

50Hz

60Hz

50Hz

60Hz

50Hz

RF Amp Type 1.5T SRFD2 RF amp

1.5T 16KW XRFD

1.5T SRFD2 RF amp

1.5T 16KW XRFD

1.5T SRFD2 RF amp

1.5T 16KW XRFD

1.5T SRFD2 RF amp

1.5T 16KW XRFD

1.5T SRFD2 RF amp

1.5T 16KW XRFD

1.5T SRFD2 RF amp

1.5T 16KW XRFD

3.0T 30KW Dual Drive RF Amp3.0T 30KW Dual Drive RF Amp3.0T 30KW Dual Drive RF Amp3.0T 30KW Dual Drive RF Amp3.0T 30KW Dual Drive RF Amp
Governing Body fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

fda2

iec

MHLW-JIS

special1

special2

ISO Vector Z 1.5T 60cm Magnet Enclosure1.5T 70cm Magnet Enclosure1.5T 70cm Magnet Enclosure1.5T 70cm Magnet Enclosure1.5T 70cm Magnet Enclosure1.5T 70cm Magnet Enclosure3.0T 60cm Magnet Enclosure3.0T 70cm Magnet Enclosure3.0T 70cm Magnet Enclosure3.0T 70cm Magnet Enclosure3.0T 70cm Magnet Enclosure
Magnet Enclosure DV EnclosureDV EnclosureGEM EnclosureArchitect/ ArtistArchitect/ ArtistArchitect/ ArtistDV EnclosureGEM EnclosureArchitect/ ArtistArchitect/ ArtistArchitect/ Artist
PAC Type PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5PAC2B/ PAC5
RF-Systems Cabinet (RFS) yesyesyesyesyesyesyesyesyesyesyes
Spectro Rfamp Type None

CPC 5118154

CPC 5118154-3

NoneNoneNoneNoneNoneNone

AMT

None

AMT

None

AMT

None

AMT

None

AMT

Scanner Channel Configuration 32

16

32

16

32329612832

16

323296128
Shim Supply Type None

NAV Shim Supply

NoneNoneNoneNoneNoneNone

NAV Shim Supply

NoneNoneNoneNone
LPCA/Port Configuration 2 P-Port on LPCA2 P-Port on LPCA1 P-Port on LPCA1 P-Port on LPCA1 P-Port on LPCA1 P-Port on LPCA2 P-Port on LPCA1 P-Port on LPCA1 P-Port on LPCA1 P-Port on LPCA1 P-Port on LPCA
HEC Type

(see HEC Troubleshooting to identify HEC type)

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

G6001EN

G6000EN

UPM Type QUPM 1MHz sampling rate

UPM 100KHz sampling rate

QUPM 1MHz sampling rate

UPM 100KHz sampling rate

QUPM 1MHz sampling rate

UPM 100KHz sampling rate

QUPM 1MHz sampling rateQUPM 1MHz sampling rateQUPM 1MHz sampling rateQUPM 1MHz sampling rate

UPM 100KHz sampling rate

QUPM 1MHz sampling rate

UPM 100KHz sampling rate

QUPM 1MHz sampling rate

UPM 100KHz sampling rate

QUPM 1MHz sampling rate

UPM 100KHz sampling rate

QUPM 1MHz sampling rate

UPM 100KHz sampling rate

Table Configuration NoneNoneNoneNoneNoneNoneNoneFlat (GEM) table

Curved table

Flat (GEM) tableFlat (GEM) tableFlat (GEM) table
SSC Type MGD

ICE

MGD

ICE

MGD

ICE

ICEICEICEMGD

ICE

MGD

ICE

ICEICEICE
Receiver Type RRxRRxRRxRRxDPP_RCVRDPP_RCVRRRxRRxRRxDPP_RCVRDPP_RCVR
Grad Processor Type NoneNoneNoneNoneNoneNoneNoneNoneNoneNoneNone